12-18
D
ELTA
/D
ELTA
XL/K
APPA
VF8
12 R
ESPIRATION
Status Messages
Message
Possible Cause
Suggested Action
RESP > #
The respiration rate is above the
upper alarm limit.
• Check the patient and treat, if necessary.
• Check the placement of electrodes. Change
their position, if necessary.
• Move the electrodes away from the source
of interference.
RESP < #
The respiration rate is below the
lower alarm limit.
• Check the patient and treat, if necessary.
• Check the placement of electrodes. Change
their position, if necessary.
RESP Out of
Range (High)
• The respiration rate exceeds
150 breaths per minute.
• The monitor may be counting
artifacts as valid breaths.
• The monitor may be counting
interference caused by faulty
equipment.
• Check the patient and treat, if necessary.
• Check the placement of electrodes. Change
their position, if necessary.
• Move the electrodes away from the source
of interference.
RESP Apnea
(neonatal or
pediatric mode
only)
No respiration has been
detected for <XX> seconds.
• Check the patient and treat, if necessary.
• Check the placement of electrodes. Change
their position, if necessary.
• Carry out a RESP relearn.
• Reset breath-detection sensitivity to manual
mode.
RESP
Coincidence
The patient’s heart rate and
respiration rate fall within 20% of
each other.
• Observe the patient and treat, if necessary.
• Check and change the electrode placement
if you receive a coincidence message until
you obtain a clear respiration signal.
RESP: Can’t
detect
coincidence
• The RESP coincidence feature
is enabled but there is
excessive ECG artifact.
• ECG leads maybe
disconnected.
• Calm the patient.
• Make sure the patient’s skin is properly
prepped.
• Isolate the patient from auxiliary equipment.
• Check the electrodes. Reapply gel or
change the electrode, if necessary.
• Inspect and replace defective cables and
wires.
• If a lead or electrode cannot be replaced,
select another lead for processing (from the
ECG menu).
RESP Signal
Saturated
RESP signal detected by
monitor has excessive baseline
shift.
• Check the patient cable and lead wires
carefully.
• Replace any cable or lead wire that is
suspect.
• Reapply gel or change the electrode.
• Check the M
ULTI
M
ED
pod and replace, if
necessary.
Содержание Infinity Delta
Страница 22: ...22 DELTA DELTA XL KAPPA VF8 INFINITY DELTA SERIES USER S GUIDE This page intentionally left blank ...
Страница 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Страница 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 102: ...4 8 DELTA DELTA XL KAPPA VF8 4 ADMISSION TRANSFER AND DISCHARGE This page intentionally left blank ...
Страница 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Страница 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Страница 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Страница 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Страница 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Страница 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Страница 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Страница 234: ...13 16 DELTA DELTA XL KAPPA VF8 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Страница 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Страница 330: ...18 16 DELTA DELTA XL KAPPA VF8 18 TRANSCUTANEOUS BLOOD GAS MONITORING This page intentionally left blank ...
Страница 358: ...20 10 DELTA DELTA XL KAPPA VF8 20 MICROSTREAM ETCO2 MONITORING This page left blank intentionally ...
Страница 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Страница 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Страница 468: ...26 30 DELTA DELTA XL KAPPA VF8 26 PULSE CONTOUR CARDIAC OUTPUT PICCO MONITORING This page intentionally left blank ...
Страница 469: ...27 Body Temperature Overview 27 2 Temperature Display 27 4 Temperature Setup 27 6 Status Messages 27 6 ...
Страница 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Страница 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Страница 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...