Infinity Acute Care System 

This Infinity Quick Reference 
Guide is not a  replacement or 
substitute for the Instructions for 
Use (IFU).  It is for informational 
purposes only

 

Software Level:  VG4

 

 

Quick Reference Guide 

 

Infinity Cockpit 

C700 / C500 
 

Содержание Infinity C500

Страница 1: ...re System This Infinity Quick Reference Guide is not a replacement or substitute for the Instructions for Use IFU It is for informational purposes only Software Level VG4 Quick Reference Guide Infinity Cockpit C700 C500 ...

Страница 2: ...rm Setup 16 18 Screen Setup Restore Profile and Views 19 20 Parameter Setup 21 ECG 21 22 RRi Impedance Respirations 22 23 SpO2 23 24 NIBP 24 Mainstream etCO2 Monitoring 25 26 12 lead ECG 26 Invasive Pressures 27 PAWP 28 Cardiac Output 28 29 Calculations 30 Warning This Quick Reference Guide is not intended to replace the Instructions for Use manual For full information concerning the performance c...

Страница 3: ...Do remove hair if necessary for proper elec trode placement 9 9Do change electrodes Q 24 48 hours or per hospital policy U Do Not utilize alcohol to dry and clean area of skin for electrode placement U Do Not place electrodes over bony promi nences or area of increased muscle move ment activity U Do Not assume electrodes are good if the tape portion is attached to the patient s skin U Do Not braid...

Страница 4: ...MANUAL ADDITIONAL INFORMATION Electrode Colors IEC AHA US Electrode Location L Yellow LA Black Left arm F Green LL Red Left leg R Red RA White Right arm N Black RL Green Right leg neutral C White V Brown Moveable chest C Gray White V Gray Brown 2nd Moveable chest Standard 6 Lead Configuration IEC AHA Standard 5 Lead Configuration IEC AHA ...

Страница 5: ...TRUCTIONS FOR USE MANUAL ADDITIONAL INFORMATION Standard 3 Lead Configuration IEC AHA Pacemaker Configuration 5 electrodes IEC AHA Warning Make sure the pacemaker detection is turned OFF for patients without pacemakers and that it is ON for patients with pacemakers Al ways keep pacemaker patients under close surveillance ...

Страница 6: ... INFINITY COCKPIT QUICK REFERENCE GUIDE 6 THIS GUIDE IS INTENDED FOR REFERENCE ONLY REFER TOINSTRUCTIONS FOR USE MANUAL ADDITIONAL INFORMATION 12 Lead Configuration ten electrodes IEC AHA Respiratory lead placement for infants ...

Страница 7: ...tral Station ICS a network re corder and a network printer 3 M540 facilitates transport of patient when undocked by main taining all patient connections and alarm settings 4 M540 when configured as wireless can send information to the ICS over a Wi Fi network TO UNDOCK M540 release for transport 1 Push locking tab toward the back of M500 2 While holding the M540 firmly with one hand press one of t...

Страница 8: ...wer On Off key 2 Select Shutdown inside the dialog window The message 3 System is shutting down appears on the screen while the 4 Infinity Medical Cockpit is being powered down ROTARY KNOB Navigates through menus and tables to define settings 1 Touch a Parameter box or Function key to display a menu Turn the Rotary knob to the left or right to increase or de crease a numeric value to scroll down a...

Страница 9: ... alarm signals corre sponding to the alarm priority If the new alarm is of lower priority than the paused alarm the new alarm condition is only represented by a visual alarm signal No acoustic alarm tones sound If Quiet Mode is Deactivated Any new alarm condition breaks through the Audio Pause period with full acoustic and visual alarm annunciation TO CALIBRATE THE C700 500 SCREEN 1 Press the rota...

Страница 10: ...TIONS FOR USE MANUAL ADDITIONAL INFORMATION Blue Header Bar Quick Access Icons Function Keys BLUE HEADER BAR Appears along the top of the screen and displays the following 1 Patient category field adult pediatric neonate 2 System data field 3 Patient name field 4 Time and date field 5 Alarm message field 6 Alarm banner field ...

Страница 11: ...t offer access to frequently used functions Quick access sym bol next to Associate tool bar ALARMS All alarms off or All alarms paused Auto set all VIEWS Show all ECG Remote view Print screen TRENDS AND DATA Trend Table Report ECG Reports Timed Waveform Report Timed Recording and Continuous Recording etc SENSOR PARAMETERS Zero all NIBP continuous Venous stasis START STANDBY Standby Discharge Priva...

Страница 12: ... for setting up screen view and Remote view PRINT SCREEN Prints the current screen view on a connect ed laser printer FREEZE WAVEFORMS Stops all waveforms for 60 seconds TRENDS AND DATA Opens the Trends Data dialog window PROCEDURES Opens the dialog window for setting up Wedge and C O SENSOR PARAME TERS Opens the dialog window for setting up each parameter NIBP START STOP When selected starts an N...

Страница 13: ...ear in blue header bar Note If you dock an M540 with a patient category that differs from the one selected on the Cockpit the patient category setting of the Cockpit changes to match the M540 ENTER PATIENTS DEMOGRAPHIC DATA FOR ADMISSION 1 Touch patient category symbol in left most field of the blue header bar on Cockpit 2 Select Name ID Birth date Admit date Physician height and weight and use ke...

Страница 14: ...pit and the M540 in standby mode 1 Select the Start Standby button on the main menu bar to display the Standby dialog window 2 Select the Start tab if it is not already selected 3 Select the Standby button The banner Standby Touch Screen to resume monitoring appears in the center of the Cockpit screen 4 To take the Cockpit out of Standby mode touch the screen to resume monitoring the vital signs T...

Страница 15: ... parameter s box and alarm message field have a yellow background and audible alarms sound when alarms are active Are non latched audio visual flashing alarms stop when situa tion corrects self but alarm status message will remain in the header bar To remove message press LOW PRIORITY ALARMS Alerts to technical issues that may compromise the ability of the system to monitor the patient for example...

Страница 16: ... alarm tones sound If Quiet Mode is Deactivated Any new alarm condition breaks through the Audio Pause period with full acoustic and visual alarm annunciation TO PRE SILENCE ALARMS FOR 2 MINUTES This function allows you to pre silence audio pause in advance potential alarm conditions before they occur 1 Press the yellow key on the Cockpit o Alarm conditions are reported visually by a corres pondin...

Страница 17: ...GS 1 Select desired parameter box to access the parameter setup page 2 Select button above the current patient value to adjust the High limit Select the button below the current patient value to adjust the Low limit 3 When button is highlighted yellow use Rotary knob to adjust the parameter limit higher or lower 4 Press Rotary knob to confirm 5 Select button below low limit to select how an alarm ...

Страница 18: ... The priority for asystole and ventricular fibrillation events cannot be changed The alarm priority high is always assigned to these categories 5 Select the corresponding setup button in the Rate column to set the rate 6 Select the corresponding setup button in the Count column to set the count 7 Select the corresponding setup button in the Archive column to set how it will be archived with alarm ...

Страница 19: ...play mode selection button to have the Main screen display only those parameters with cables connected to the monitor OR 3 Select Manual display mode to have the Main screen show all parameters even those not connected 4 Adjust position and number of waveforms and parameter boxes turn on Split screen and setup and turn on Mini trends display AUTO VIEW TOOLBAR 1 On Auto view toolbar at bottom of sc...

Страница 20: ...rofile has now been restored Note Patient category does not change automatically TO SELECT A DIFFERENT VIEW A View refers to the layout and content of the screen Each profile has a pre defined view associated with it Each profile can also have alternate views associated with it 1 Select Views key on main menu bar 2 Select name of desired view Auto in the name indicates a plug and play concept wher...

Страница 21: ...alog box ECG TO ADJUST PULSE TONE ON OFF 1 Select Heart rate parameter box 2 At Pulse tone volume select button 3 Use rotary knob to dial and select desired setting 4 Click rotary knob to confirm TO CHANGE LEADS 1 Select Heart rate parameter box 2 At Leads select 1st 2nd or 3rd button to choose which dis played channel on screen to change 3 Make selection from available leads 4 Click rotary knob t...

Страница 22: ...ameter box 2 Select Settings 2 tab 3 Select Normal or Low QRS Threshold 4 Click rotary knob to confirm Note Low detects QRS complexes of low amplitude TO SHOW ALL LEADS AND OR PRINT ECG REPORT 5 Select Heart rate parameter box 6 Select Show all tab 7 Select ECG report Print for non diagnostic report 8 Select Rest ECG report Print for diagnostic report o Diagnostic report requires specific software...

Страница 23: ... RRi waveform regularly to reduce artifact and optimize breath count o Manual The monitor detects breath based on the size of the waveform After turning on Manual mode adjust waveform size Manual mode is intended for adult or pe diatric patients whose breathing patterns show excessive variation or for neonates whose breathing rhythms tend to be irregular and whose signals may not be reliably de te...

Страница 24: ...select ECG or SpO2 Note Refer to M540 Instructions for Use manual for further device specific information Masimo or Nellcor NIBP Non Invasive Blood Pressure TO START A NIBP SINGLE MEASUREMENT 1 Press the NIBP Start Stop fixed key o Pressing the key again stops the measurement o NIBP inflation modes Adult pediatric and neonatal can be selected in NIBP parameter box and Settings tab TO SET NIBP INTE...

Страница 25: ...ainstream etCO2 Monitoring CONNECTING THE CO2 SENSOR 1 Connect yellow end of CO2 sensor cable to yellow CO2 con nector on M540 2 Insert correct airway adapter for patient between endo tracheal tube adapter and the ventilator Y piece 3 Snap the CO2 mainstream sensor firmly onto the airway adapter and make sure the cable is directed away from the pa tient Note Always position the sensor windows of t...

Страница 26: ...irway adapter before attempting to zero o Do not breathe on sensor during calibration o Make sure adapter is free from secretions and moisture 2 Select CO2 parameter box 3 Select Zero o Zero failed or Zero accepted will be displayed in blue header bar TO SET RRc APNEA TIME 1 Select CO2 parameter box 2 Select RRc Apnea time secs 3 Dial rotary knob to select desired setting 4 Click rotary knob to co...

Страница 27: ...ed and patient is admit ted at ICS with Name Height Weight and Birth date TO OBTAIN A DIAGNOSTIC 12 LEAD REST ECG REPORT 1 Select System setup on the main menu bar 2 Select the Recordings Reports tab 3 Select Rest ECG report 4 Enter patient gender and race condition and medications 5 Press Rest ECG report OR 1 Press Menu function key 2 Press Rest ECG report Note Confirm all leads are free from art...

Страница 28: ...t Sensor parameters from main menu bar 6 Select IBP tab to access IBP page 7 Select desired IBP tab along right side of IBP page 8 Select Label 9 Dial rotary knob to select desired setting 10 Click rotary knob to confirm TO ZERO ALL TRANSDUCERS 1 Level transducer stopcocks appropriately for parameters being zeroed 2 Close the transducer stopcocks to the patient and open them to air 3 Select Zero a...

Страница 29: ...tons o Freeze Adjust button to alter the PWP value manually o Save wedge button to save the new value o Cancel wedge to cancel the measurement Cardiac Output BEFORE INJECTING FOR A CARDIAC OUTPUT 1 Dräger recommends pre filled syringes or a closed injectate system placed in an ice bath maintained between 0 C 32 F and 5 C 41 F The accuracy of measurements done with the thermodilution method increas...

Страница 30: ...M A CARDIAC OUTPUT MEASUREMENT 1 Select C O parameter box o If necessary select Mode Auto or Manual 2 Select Start C O OR 3 Select the Procedures button from the main menu bar 4 Select the C O tab if necessary 5 Select Start C O o In Auto mode the message Inject when ready appears in the Procedures C O a Blood temperature must be stable from previous in jectate solution otherwise will give error m...

Страница 31: ...ar 2 Select Calculations tab o Touch Capture values All necessary data available to the monitor will be captured o To change or manually enter data touch the value or values to be changed a Using adjacent on screen keypad or keyboard touch to enter data o Click rotary knob to confirm 3 If active lab results can be added by touching Capture labs 4 Touch Calculate Results 5 Touch Results tab on righ...

Страница 32: ...raeger Medical Canada Inc 2425 Skymark Ave Unit1 Mississauga ON L4W 4Y6 Tel Toll free Fax 1 905 212 6600 1 866 343 2273 1 905 Manufacturer USA Draeger Medical Inc 3135 Quarry Road Telford PA 18969 1042 USA Draeger Medical Systems Inc Telford PA 18969 USA The quality management sys tem at Draeger Medial Sys tems is certified according to Tel Toll free Fax 1 215 721 5400 1 800 437 2437 1 215 723 593...