Страница 1: ...Supplement to the Fabius GS Fabius GS premium Fabius Tiro and Fabius OS Operator s Instruction Manuals for the Optional Heated Breathing System Supplement 9038278 Rev 07 ...

Страница 2: ...This page intentionally left blank Supplement 9038278 Rev 07 Fabius GS GS premium Tiro OS Supplement ...

Страница 3: ...Document 3 Heated Breathing System Optional 3 Symbol Definition 3 General Warnings and Cautions 4 Installation of HBSPS Fabius GS GS premium OS 5 Installation of HBSPS Fabius Tiro 7 Operation of HBSPS 12 Switching Off HBSPS 14 Preparing the Compact Breathing System for Maintenance 14 HBSPS Technical Data 15 Daily Checkout 16 ...

Страница 4: ...This page intentionally left blank 2 Supplement 9038278 Rev 07 Fabius GS GS premium Tiro OS Supplement ...

Страница 5: ...m Optional The Fabius can be configured with a heated breathing system to reduce condensation of moisture in the system Symbol Definition The following symbols appear on the Heated Breathing System Power Supply HBSPS Optional Caution Refer to accompanying documents before operating equipment Caution Risk of electric shock do not remove cover Refer servicing to a DrägerService representative Degree...

Страница 6: ...When the moisture content falls below a specified minimum level the following undesirable reactions can occur regardless of the type of CO2 absorbent and the anesthetic agent used e g Halothane Enflurane Isoflurane Sevoflurane or Desflurane reduced CO2 absorption formation of CO absorption and or decomposition of the inhalation anesthetic agent increased heat generation in the absorber leading to ...

Страница 7: ... the left rear GCX rail of the Fabius by sliding the mounting plate of the power supply through the opening of the rail 1 and then sliding it upward Tighten the knob 2 against the rail when the power supply is in the desired position 2 Route the power cable 3 for the heater below and behind the power supply and then up to the rear of the COSY arm Push the cable into the cable management cavity 4 o...

Страница 8: ...ighten the knob 1 in Figure 2 against the rail when the power supply is in the desired position 2 Route the power cable for the heater through the trough 2 in Figure 2 on the rear of the cabinet At the right hand end of the trough route the cable into the cable management cavity 1 in Figure 3 on the bottom of the COSY arm Figure 3 HBSPS Installation on Machine with Right Mount COSY Step 2 3 Plug t...

Страница 9: ...ft rear GCX rail of the Fabius Tiro by sliding the mounting plate of the power supply through the opening of the rail 1 and then sliding it upward Tighten the knob 2 against the rail when the power supply is in the desired position 2 Route the power cable 3 for the heater below and behind the power supply and then up to the rear of the COSY arm Push the cable into the cable management cavity 4 on ...

Страница 10: ...ighten the knob 1 in Figure 5 against the rail when the power supply is in the desired position 2 Route the power cable for the heater through the trough 2 in Figure 5 on the rear of the cabinet At the right hand end of the trough route the cable into the cable management cavity 1 in Figure 6 on the bottom of the COSY arm Figure 6 HBSPS Installation on Machine with Right Mount COSY Step 2 Trolley ...

Страница 11: ...n Figure 7 from the side of the power supply by removing the two screws 2 in Figure 7 Figure 8 HBSPS Installation on Wall Mounted Machine Step 2 2 Install the Draeger clamp onto the rear side of the power supply box using the two screws and lockwashers included with the clamp assembly Figure 8 Figure 9 HBSPS Installation on Wall Mounted Machine Step 3 3 Clamp the power supply onto the rear area of...

Страница 12: ...d the power supply and then up to the rear of the COSY arm 1 in Figure 10 Extra cable can be coiled around the clamp Push the cable into the cable management cavity on the bottom of the COSY arm 1 in Figure 11 5 Plug the connector at the end of the cable into the mating connector on the rear underside of the COSY 2 in Figure 11 Orient the connector so that the red mark on the connector plug aligns...

Страница 13: ...hine route the power cable for the heater through the trough on the rear of the cabinet 1 in Figure 12 At the right hand end of the trough route the cable into the cable management cavity on the bottom of the COSY arm 2 in Figure 12 5 Plug the connector at the end of the cable into the mating connector on the rear underside of the COSY 3 in Figure 12 Orient the connector so that the red mark on th...

Страница 14: ...ater be turned off when not in use Warning When the moisture content falls below a specified minimum level the following undesirable reactions can occur regardless of the type of CO2 absorbent and the anesthetic agent used e g Halothane Enflurane Isoflurane Sevoflurane or Desflurane reduced CO2 absorption formation of CO absorption and or decomposition of the inhalation anesthetic agent increased ...

Страница 15: ...17 is plugged into a functioning wall outlet Figure 16 HBSPS Connector HEATER POWER ON LED Is Not Illuminated If the HEATER POWER ON LED 2 in Figure 15 is not illuminated when the HEATER POWER switch 1 in Figure 15 is in the ON position and the MAINS POWER LED 1 in Figure 17 is illuminated check the fuse 3 in Figure 15 If it is blown replace it with the properly rated fuse Figure 17 HBSPS Back MAI...

Страница 16: ... System for Disinfection and Cleaning 1 Leave the Dräger Vapor s on the machine 2 Remove all breathing hoses 3 Remove the breathing bag 4 Remove the ventilation hose 5 Remove the fresh gas hose from the breathing system 6 Remove the anesthetic scavenging hose 7 Detach the APL bypass and the Peep Pmax lines from the breathing system and from the side of the machine 8 Remove the flow sensor cable 9 ...

Страница 17: ... Rev 07 15 Disinfection Cleaning Autoclaving The same information applies to the breathing system with heater as that for the breathing system contained in the Fabius Operator s Instruction Manual Cleaning disinfection of the mains adapter should be performed as described in the Fabius Operator s Instruction Manual under Machine exterior ...

Страница 18: ... 60127 2 V 1x F1 60AH 250V IEC 60127 2 V Electromagnetic Compatibility EMC Conforming to EN 60601 1 2 IEC 60601 1 2 and ANSI AAMI IEC 60601 1 2 Electrical Safety Conformance IEC 60601 1 CAN CSA C22 2 No 601 1 M90 UL 60601 1 Environmental Temperature Operating 10 C to 35 C Storage 10 C to 70 C Humidity Operating 20 to 80 non condensing Storage 10 to 90 non condensing Barometric Pressure Operating 7...

Страница 19: ...that the MAINS POWER LED is illuminated Ensure that the HEATER POWER switch is in the ON position Ensure that the HEATER POWER ON LED is illuminated Ensure that the underside of the COSY is warm approximately 35 40 C after 30 minutes have elapsed since the HEATER POWER ON LED was illuminated Fabius GS GS premium Tiro OS Supplement Supplement 9038278 Rev 07 17 ...

Страница 20: ...ibuted by Draeger Medical Inc 3135 Quarry Road Telford PA 18969 U S A 215 721 5400 800 4DRAGER 800 437 2437 FAX 215 723 5935 http www draeger com 9038278 enUS Dräger Medical GmbH Rev 08 Dräger reserves the right to make modifications to the medical device without prior notice As of 2015 08 Dräger Medical GmbH changes to Drägerwerk AG Co KGaA ...