Instructions for use BiliLux
13
Safety-related information
2.3.9
Safe connection with other electrical equipment
Unapproved electrical connections can lead to patient injury or device failure.
►
Electrical connections to equipment not listed in these instructions for use or
these assembly instructions must only be made when approved by each
respective manufacturer.
►
Connecting electrical equipment to a power socket strip effectively leads to
creating a medical electrical system. The result can be a reduced level of safety.
Comply with the requirements of IEC 60601-1 when constructing a medical
electrical system.
2.4
Device compatibility safety
The use of incompatible devices may adversely affect the functional integrity of the
product. Personal injury and property damage may occur as a consequence.
►
Ensure that other medical devices that are used in conjunction with the patient
or that form a system with this device are in compliance with the applicable
medical device requirements.
►
Do not use the device adjacent to or stacked with other equipment unless
verification of normal operation in such a configuration can be achieved.
►
If other devices are used with the BiliLux, refer to their manufacturer’s
instructions for use.
2.5
Safety instructions for accessories
2.5.1
Accessories
The use of incompatible accessories may adversely affect the functional integrity of
the product. Personal injury and property damage may occur as a consequence.
►
Dräger has tested the compatibility only of accessories listed in the current list of
accessories.
►
Use only compatible accessories. The accessories that are compatible with this
product are listed in the list of accessories supplied with the product.
2.5.2
Reusing or reprocessing accessories
Reusing, reprocessing, or sterilizing disposable products can lead to a failure of the
medical device and cause injury to the patient.
►
Do not reuse, reprocess, or sterilize disposable products. Disposable products
were designed, tested, and manufactured for one-time use only.
►
Do not use packaged disposable products if the package has been
compromised.
Reusable accessories have a limited service life. Using accessories beyond their
service life can lead to a failure of the medical device and cause injury to the
patient.
►
If the accessory shows external signs of wear such as cracks, deformations,
discoloration, or peeling, exchange the accessory.
Содержание BiliLux
Страница 2: ...2 Instructions for use BiliLux This page has been left blank intentionally ...
Страница 6: ...6 Instructions for use BiliLux This page has been left blank intentionally ...
Страница 8: ...8 Instructions for use BiliLux This page has been left blank intentionally ...
Страница 18: ...18 Instructions for use BiliLux This page has been left blank intentionally ...
Страница 26: ...26 Instructions for use BiliLux This page has been left blank intentionally ...
Страница 30: ...30 Instructions for use BiliLux This page has been left blank intentionally ...
Страница 36: ...36 Instructions for use BiliLux This page has been left blank intentionally ...
Страница 54: ...54 Instructions for use BiliLux This page has been left blank intentionally ...
Страница 58: ...58 Instructions for use BiliLux This page has been left blank intentionally ...
Страница 64: ...64 Instructions for use BiliLux This page has been left blank intentionally ...
Страница 70: ...70 Instructions for use BiliLux This page has been left blank intentionally ...
Страница 84: ...84 Instructions for use BiliLux This page has been left blank intentionally ...
Страница 88: ...88 Instructions for use BiliLux This page has been left blank intentionally ...
Страница 89: ...Instructions for use BiliLux 89 This page has been left blank intentionally ...