Instructions for use Babyleo TN500 SW 1.0n
75
Operation
6.7.1.1
Temperature adjustment during transition
The table shows how the temperature is adjusted during the transition in air
temperature mode, in skin temperature mode, and in manual mode:
6.7.1.2
Interrupting incubator operation
If the hood was opened for less than 2 hours, the device continues operating with
the most recent air temperature settings for incubator operation once the hood is
closed. The settings for oxygen (option) and humidity are also retained.
Example:
The hood is opened for a few minutes to check the position of the
ventilation mask.
Mode
Transition
Temperature
Air temperature
mode
The radiant warmer power
corresponds to the air tem-
perature that was set with the
hood closed. The
Check set-
tings
alarm indicates that the
user must confirm or adjust
this setting.
Skin tempera-
ture mode
The device adopts the skin
temperature value that was
set with the hood closed.
The device adopts the skin
temperature value that was
set with the hood open.
Manual mode
The air temperature corre-
sponds to the radiant warmer
power that was set with the
hood open. The
Check set-
tings
alarm indicates that the
user must confirm or adjust
this setting.
Содержание Babyleo TN500
Страница 2: ...2 Instructions for use Babyleo TN500 SW 1 0n This page has been left blank intentionally...
Страница 212: ...212 Instructions for use Babyleo TN500 SW 1 0n This page has been left blank intentionally...
Страница 214: ...214 Instructions for use Babyleo TN500 SW 1 0n This page has been left blank intentionally...
Страница 221: ...Instructions for use Babyleo TN500 SW 1 0n 221 This page has been left blank intentionally...