Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Operation
4. Adjust the patient data and ventilation settings as soon as possible.
7.2.1
Loading the patient data
There are two possibilities for loading the patient data:
– Starting a new case (1)
– Continuing a case (2)
Depending on the selected patient category, different patient data (e.g., weight,
age) are displayed in area (3).
7.2.1.1
Starting a new case
►
Touch the button for the desired patient category (1).
For new patients, the device uses defined start settings for ventilation settings and
alarm limits, see page 166. The set value for
Ti
is automatically set based on
RR
in
such a way that the resulting
I:E
ratio is 1:1 for neonates and 1:2 for all other patient
categories.
7.2.1.2
Continuing a case after an interruption
►
Touch the
Continue case
button (2).
The device continues to use the ventilation settings and alarm limits that were set at
the start of the case.
This function is only available if a case has previously been started and then
interrupted.
3529
7
3
1
2
Start
Select patient and check data
Select mode