Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Getting started
6.5 Emergency
start-up
6.5.1
Electronically controlled gas mixer
1. Adjust the APL valve.
2. Set the O
2
switch (3) to the
Add. O
2
position.
3. Open the flow control valve (2) and set the desired O
2
flow. If necessary, press
the
O
2
+
key (4) to quickly fill the breathing bag.
4. Monitor the set flow on the O
2
flowmeter (1).
5. Set the anesthetic gas concentration on the vaporizer.
6. Ventilate the patient manually.
7. Turn on the device.
8. As soon as the
Standby
page is displayed, start the therapy, see page 93.
9. Set the O
2
switch (3) to
Aux. O
2
.
10.Close the flow control valve (2).
6.5.2
Mechanically controlled gas mixer
1. Adjust the APL valve.
2. Open the flow control valve (1) and set the desired O
2
flow. If necessary, press
the
O
2
+
key (3) to quickly fill the breathing bag.
3. Monitor the flow on the total flow tube (2).
4. Set the anesthetic gas concentration on the vaporizer.
5. Ventilate the patient manually.
6. Turn on the device.
7. As soon as the
Standby
page is displayed, start the therapy, see page 93.
35041
3
2
1
4
35054
1
3
2