Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Assembly and preparation
4. Check the flow indicator.
If the flow indicator (4) is floating in the normal range (5), the anesthetic gas
scavenging system is functional.
If the flow indicator is floating in the restricted range (6), certain fresh-gas flows
should not be exceeded, see "Anesthetic gas receiving system" in chapter
"Technical data".
4.6.2
Passive anesthetic gas scavenging
1.
WARNING
Risk of negative pressure
If a passive AGS is connected to an active anesthetic gas scavenging system, a
negative pressure may arise in the patient’s lungs.
►
Connecting a passive AGS to an active anesthetic gas scavenging system is
not allowed.
WARNING
Risk of overpressure
If the overpressure valve in the passive AGS or the scavenging hose is blocked,
overpressure in the breathing circuit and in the patient’s lungs will occur.
►
Only connect the passive AGS to kink-proof and pressure-tight scavenging
hoses.
WARNING
Risk of ambient air contamination
The ambient air will be contaminated if the passive AGS feeds excess
anesthetic gas to a ventilation system with circulating air.
►
Use the passive AGS only with ventilation systems that work without
circulating air.
40960
5
6
4