58
Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Assembly and preparation
4.3.5
Connecting other devices to the auxiliary power sockets
An illustration of the auxiliary power sockets can be found on page 28.
1. Connect the power cable of the other device to an auxiliary power socket.
Make sure that the maximum current consumption of the other devices does not
exceed the permissible value.
WARNING
Risk of fire
Components such as power supply units that heat up are unable to cool down in
enclosed storage locations and may cause a fire.
►
Do not keep components that heat up in the drawers or in the storage
compartment at the rear of the device.
4.3.6 Establishing
potential equalization
Differences in electrical potential between devices can be reduced by potential
equalization.
Potential equalization does not replace the connection with protective ground
conductor.
During operation, the potential equalization connector must be readily accessible
and the potential equalization cable must be removable without tools.
4.3.6.1
Connecting the potential equalization cable
– Connect the potential equalization cable to the potential equalization pin on the
device (see "Connectors", page 25).
– Connect the potential equalization cable to a potential equalization connector of
the hospital (e.g., wall, ceiling supply unit, operating table).
4.3.7
Connecting devices to the data interfaces
This device is equipped with data interfaces such as LAN and RS232. These
interfaces can be used to set up an IT network in accordance with IEC 60601-1.
4.3.7.1
Establishing a data connection
►
WARNING
Risk of overloading the network
The following hazardous situations may occur if the network does not possess
the required characteristics:
►
Exported patient-related data (age, weight, height) and therapy-related data
may be intercepted, falsified, or damaged.
►
An overload of the device due to high network load (e.g., caused by denial-
of-service attacks) may lead to a shut-down of the device’s network port. The
network port will not be available again until the device is restarted.
Connect the device to a network or a computer.
An illustration of the ports can be found on page 25.