Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Overview
3.2.12.2 Backup
manual
mode
In various technical fault situations, the backup manual mode enables a direct
changeover to manual ventilation in order to continue the therapy.
Backup manual mode with an electronically controlled gas mixer
1. Open the flap (1).
2. Activate the backup manual switch (2). Follow the instructions on the product
label (3) or the screen (4).
3. Set the O
2
switch to
Add. O
2
.
4. Open the flow control valve on the emergency O
2
delivery and set an adequate
O
2
flow. The set O
2
flow constitutes the total fresh-gas flow.
5. Ventilate the patient manually.
a. During mechanical ventilation or in the
Man/Spon
ventilation mode:
Ventilate manually with the breathing bag.
b. When using the external fresh-gas outlet:
Ventilate manually with the breathing bag on the non-rebreathing system.
6. Check the vaporizer setting.