Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
37
Overview
If the connected network offers an NTP service, the time on the device can be
synchronized with the time on the NTP server.
For further information see: "Connections to IT networks", page 294.
3.2.11.4
Support of Infinity ID accessories
– Replacement interval monitoring
– Anti-interchange security for breathing hoses
For further information see: "Support of Infinity ID accessories", page 312.
3.2.12 Safety
functions
3.2.12.1 Emergency
O
2
delivery (electronically controlled gas mixer)
CAUTION
Risk of increased anesthetic agent delivery
When the emergency O
2
delivery (
Add. O
2
) is in use, anesthetic agent continues to
be delivered into the breathing system in accordance with the vaporizer setting.
When the emergency O
2
delivery is used during low-flow anesthesia or minimal-
flow anesthesia, an increased quantity of anesthetic agent may enter the breathing
system. This may lead to an increased anesthetic gas concentration.
►
Carefully monitor the gas mixture.
1. Check the vaporizer setting.
2. Set the O
2
switch (1) upwards to the
Add. O
2
position.
3. Open the flow control valve (2) on the O
2
flowmeter and set the desired flow.
This O
2
flow flows through the vaporizer.