Reprocessing
13.9.2
Disassembling the components
13.9.2.1 CLIC
adapter (option)
1. Screw off the CLIC adapter (1).
2. Press the release button (2). The holder (3) flips open.
3. Push the cleaning plate (4) fully into the adapter.
4. Close the holder (3) so that it engages.
13.9.2.2
Flow sensors and ports
CAUTION
Risk of failure of flow measurement
Improper reprocessing and soiling, such as residues or particles, can damage the
flow sensor.
►
No machine cleaning or disinfection
►
No plasma sterilization or radiation sterilization
►
No water jets, compressed air, brushes or the like
►
No ultrasonic bath
►
No hot-steam sterilization with Spirolog and Infinity ID flow sensors
►
Clean and disinfect the flow sensor in accordance with the corresponding
instructions for use.
►
For disinfecting the flow sensor use only clean disinfectant solutions.