Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Reprocessing
Remove the breathing system (2) vertically upwards from the breathing system
mount.
13.8.7
Removing the piston diaphragm
►
Remove the piston diaphragm (1) from the ventilator. If necessary, remove any
residual moisture present.
13.8.8 Disinfecting
the
breathing system mount
Note the following information:
– Do not use alcohol-based disinfectants.
– Remove any residues.
– Ensure lint-free cleaning and disinfection.
– Do not allow fluid to penetrate.
Prerequisite:
– The breathing system cover has been removed.
– The breathing system has been removed.