Troubleshooting
When the breathing bag is removed, the missing fresh-gas volume will
automatically be filled by ambient air. Nevertheless, the
Fresh gas low or leakage
alarm is likely to be triggered.
12.3.4 After
the
central supply is restored
1. Connect the compressed gas hoses to the terminal units.
2. Close the gas cylinder valve on the corresponding gas cylinder again.
On devices that are equipped with Advanced Cylinder Support, the gas cylinder
valves can remain open.
12.4
Failure of fresh-gas delivery (electronically controlled
gas mixture)
If the fresh-gas delivery has failed, the emergency O
2
delivery can be used to
deliver oxygen and anesthetic agent. The current ventilation mode and the fresh-
gas deficiency detection remain active.
In the event of a fault, illustrations and instructions showing how to start the
emergency O
2
delivery are displayed in areas (1) and (2).
The emergency O
2
delivery is started as follows:
WARNING
Risk of patient injury
If the gas mixer fails, no fresh gas is delivered.
►
Check the vaporizer setting.
►
Supply the patient with O
2
.
►
Use the emergency O
2
delivery.
1. Set the O
2
switch upwards to the
Add. O
2
position. (Follow the illustration on the
screen.)
The
Internal FG flow failure
alarm will then be automatically downgraded.
2. Open the flow control valve on the O
2
flowmeter. Set the desired flow. This O
2
flow flows through the vaporizer.
36243
1
2
Internal FG flow failure