180
Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Configuration
11.3.6.7 Vertical
tab "Infinity ID"
11.3.6.8
Vertical tab "Service"
11.3.7
Resetting the start settings
Certain pages in the
System setup
dialog have a button for resetting the respective
start settings to the factory defaults.
11.3.7.1
Resetting the changes on a page
1. Open the appropriate tab.
2. Touch the
Factory defaults
button and confirm.
11.3.8 Resetting
the
consumptions
The gas consumption can be reset in
Standby
>
System setup
>
System status
>
Consumption
.
►
Touch the
Reset data
button and confirm.
Headline/
Parameter
Setting range
Description
Monitoring of Infinity ID accessories
Breathing circuit
On
; Off
Activates or deactivates the Infinity ID
functionality.
On:
– Generates a message when the
maximum period of use is exceeded
– Generates a message when
Infinity ID breathing hoses are incor-
rectly connected
Off:
– Messages are suppressed.
Water trap
On
; Off
Flow sensors
On
; Off
CO
2
absorber
On
; Off
Exchange interval [days]
Breathing circuit
Off; 2 to 9
2
Specifies the replacement intervals for
Infinity ID accessories.
Water trap
Off;
28
Flow sensors
Off; 1 to 180
90
CO
2
absorber
Off; 1 to 28
7
Headline/
Parameter
Setting range
Description
Service access code
The following functions are available
after the appropriate credentials have
been entered:
– Access to the service dialog