![Dräger Atlan A300 Скачать руководство пользователя страница 165](http://html.mh-extra.com/html/drager/atlan-a300/atlan-a300_instructions-for-use-manual_2526728165.webp)
Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
165
Configuration
11 Configuration
11.1 Device
settings
Some device functions are available as an option and consequently are only
available on appropriately equipped devices.
11.1.1 Factory
defaults
Dräger delivers the device with factory defaults that are used when starting the
device for the first time. Service personnel can reset the device to the factory
defaults.
11.1.2 Start
settings
Start settings take effect after every restart of the device or when starting a new
case (touching one of the
New adult
,
New pediatric
or
New neonate
buttons). The
start settings can be adjusted after the configuration password is entered.
If required, the device can be delivered with start settings that may differ from the
factory defaults.
11.1.3 User-specific
settings
User-specific settings can be adjusted by the user without a configuration
password. The settings take effect immediately but are discarded at the latest after
a device restart.
11.2
Setting the date and time
The device can adopt the time from a network or from a device connected via
MEDIBUS. The time synchronization takes place shortly after switch-on and at
regular intervals thereafter.
If time synchronization is not set, the time can be changed manually in 2 ways:
►
Touch the field (1).
36
019