Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
151
Ending operation
9 Ending
operation
9.1
At the end of the OR day
Dräger recommends shutting down the device during longer periods of non-use
such as overnight or on weekends. This can lower power consumption and prolong
the life span of the medical device without negatively influencing device availability.
1. Make sure that all flow control valves are closed.
2. Press the key.
3. Follow the instructions on the screen.
9.2
Storing the device
Proceed as follows when the device is to be disconnected from the mains power
supply for longer than 2 weeks:
1. Set the main switch to the
0
position.
2. Disconnect the power plug. This will prevent deep discharge of the battery.
9.3 Disconnecting
the mains power supply
When the device is disconnected from the mains power supply, the internal battery
takes over the function of supplying power. The status display remains active.
Follow the instructions for storing the device.
►
Disconnect the power plug.