background image

4

Dräger Alcotest 6510

For your safety

1

For your safety

Follow the instructions for use

Any use of the device requires understanding and observation of these 
instructions.
The device is only to be used for the purposes specified herein.

Maintenance

Maintenance must only be performed by experts.
All repairs are carried out byDräger or the Medical Bureau of Road Sa-
fety (MBRS).
Only authentic spare parts may be used for maintenance.

Liability for proper function or damage

The device is not to be serviced or repaired by any person not employ-
ed or authorised by the MBRS.
Dräger cannot be held responsible for damage caused by non-
compliance with the recommendations given above.
The warranty and liability provisions of the terms of sale and delivery 
of Dräger are likewise not modified by the recommendations given 
above.

Dräger Safety AG & Co. KGaA

2

Intended use

The Alcotest

®

 6510

1)

 is an apparatus for indicating the presence of al-

cohol in the breath. The Dual Limit Dräger Alcotest 6510 device was 
approved by the MBRS on 9th December 2010. For mobile use with 
batteries.

2.1

Preparation

The Alcotest 6510 is supplied calibrated to the MBRS by the manufac-
turer. If the calibration date exceeds 6 months arrange to have the de-
vice calibrated by the MBRS. The date of the last calibration can be 
found in the menu (page 11).

1)

Alcotest is a registered trademark of Dräger.

00123840.eps

OK

Alcotest 6510

M

Содержание Alcotest 6510

Страница 1: ...Dr ger Alcotest 6510 Dual Limit Breath Alcohol Screening Device Approved by the Medical Bureau of Road Safety MBRS Instructions for Use i...

Страница 2: ......

Страница 3: ...4 5 Switching on and off 8 5 Performing the measurement 8 5 1 Requirements of the person to be tested 8 5 2 Automatic measurement 9 5 3 Error in delivering the breath sample 10 5 4 Removing the mouth...

Страница 4: ...sible for damage caused by non compliance with the recommendations given above The warranty and liability provisions of the terms of sale and delivery of Dr ger are likewise not modified by the recomm...

Страница 5: ...election of menu functions 3 OK button Confirms entries or selected functions Button to switch device on and off 4 Menu button DOWN MENU Entering menu functions Selection of menu functions 5 Socket no...

Страница 6: ...ompartment and remove cover 2 Observe correct polarity when inserting the batteries 3 Close battery cover 4 2 Batteries to be used 2 each 1 5 V Type Alkaline Mignon LR6 AA or rechargeable battery 2 1...

Страница 7: ...ping until the mouthpiece is correctly fitted to the device 2 01123840 eps With the grooved markings facing up press the mouth piece into mouthpiece receptacle until a click sound confirms correct en...

Страница 8: ...ions z A new mouthpiece is to be used for every person tested z Do not apply too high alcohol concentrations on to sensor e g by rinsing the mouth with high grade alcohols shortly before the measureme...

Страница 9: ...utton to confirm z 02623840 eps READY Test No 00140 After approximately 6 seconds READY is displayed and a short audible signal is given z At the same time the current test number is indicated A green...

Страница 10: ...d Use the M and buttons to choose person and the OK button to confirm The device is ready for measurement after 4 seconds If the subject exhales incorrectly e g by abruptly stopping or sucking at the...

Страница 11: ...on is highlighted z Activate the selected function by pressing OK button z Selecting Back to measurement and pressing the OK button brings the device into the measuring mode If no button is pressed wi...

Страница 12: ...may show the recovery time counting down to zero 8 Scroll up to Back to measurement 9 Press the OK button The device returns to the Ready mode 7 Maintenance The operator may change the batteries as re...

Страница 13: ...teries z Gently press onto the lock of the battery compartment and remove cover z Remove used batteries and in sert new ones consider cor rect polarity z Close battery cover 7 5 1 Required batteries...

Страница 14: ...es can be disposed of using special collecting and recycling facilities However as this device has not been registered for household usage it must not be disposed of through these means The device can...

Страница 15: ...ds Re select menu Device cannot be switched on Batteries discharged Replace batteries Insuffient volume Blow interruption Subject not blowing hard or evenly enough Instruct the subject to blow in the...

Страница 16: ......

Страница 17: ......

Страница 18: ......

Страница 19: ......

Страница 20: ...754 500 Dr ger Safety AG Co KGaA Edition 04 October 2011 Edition 01 April 2005 Subject to alteration Dr ger Safety AG Co KGaA Revalstra e 1 23560 L beck Germany Tel 49 451 882 0 Fax 49 451 882 20 80 w...

Отзывы: