Intended use

The Dräger-Tube is used to detect any potassic and/or sodium cyanide aerosols in the 
air or in technical gases at the time of measurement. The Dräger-Tube only is intended 
for selective measurements.

Reaction principle

2 KCN + H

2

SO

4

 

 2 HCN + K

2

SO

4

2 HCN + HgCl

2

 

 2HCI + Hg(CN)

2

HCI + methyl red 

 red reaction product

Measurement and evaluation

1.

Ensure the measuring site has good lighting.

2.

Break off both ends of the Dräger-Tube with the Dräger-Tube opener. (Be careful 
when opening, glass splinters may fly around.)

3.

Insert the Dräger-Tube into the Dräger-Tube pump. The arrow has to point toward 
the Dräger-Tube pump

 (4) 

because otherwise measurement is impossible.

4.

Suck air or gas sample through the tube (10 strokes).

5.

Break open the interior glass ampule by bending the Dräger-Tube between the two 
black points 

(1) 

by approx. 45°. Verify if the interior glass ampule really has been 

broken open.

6.

Shake the Tube so that the ampule liquid completely adheres to the white filter and 
separation layer and carefully perform 2 pump strokes in cyanide free air. The 
indication layer must not become wet.

7.

Check the overall length of the discoloration.

8.

Multiply the value by factor F to allow for the actual atmospheric pressure.

9.

Remove the Dräger-Tube and suck in 3 to 4 strokes of clean air to flush the pump 
with air after use.

Technical data

Ambient operating conditions

Cross sensitivities

Acid gases are indicated with a different sensitivity.

Free hydrogen cyanide already is indicated before breaking open the test tube.

Due to hydrolysis, a certain part of the cyanides already may have reacted with the 
CO

2

 in the air.

It is impossible to measure cyanides in the presence of phosphine.

Disposal

The Dräger-Tube packaging lists information about the order number, shelf life, storing 
temperature limits, and the serial number. Dispose of Dräger-Tubes in accordance with 
local state and federal regulations. Keep out of reach of children and unauthorized 
persons.

Questions

If you have any questions or doubts about how to use Dräger-Tubes properly, please 
contact your Dräger Safety branch or representative office. For customers in the U.S.: 
Call Dräger Safety Inc. toll free at 1-888-794-3806.

 

Uso previsto

El tubo Dräger sirve para determinar la presencia de aerosoles de potasio y/o cianuro 
de sodio en el aire, o bien gases técnicos, en el momento de la medición. El tubo Dräger 
está indicado exclusivamente para la medición puntual.

Principio reactivo 

2 KCN + H

2

SO

4

 

 2 HCN + K

2

SO

4

2 HCN + HgCl

2

 

 2HCI + Hg(CN)

2

HCI + rojo de metilo 

 producto de reacción rojo

Realización y evaluación de la medición

1

Asegurar que el entorno de medición esté bien iluminado.

2.

Romper las dos puntas del tubo Dräger con el abridor de tubos. (Precaución al 
abrirlos, pueden saltar esquirlas).

3.

Colocar el tubo Dräger en la bomba para tubos Dräger. La flecha debe apuntar hacia 
la bomba para tubos Dräger 

(4)

, puesto que de lo contrario no es posible ninguna 

medición.

4.

Aspirar la muestra de aire o gas a través del tubo (10 emboladas).

5.

Romper la ampolla de vidrio interna doblando el tubo Dräger unos 45° entre los dos 
puntos negros 

(1)

. Comprobar que la ampolla de vidrio interna se haya roto realmente.

enUS

For your safety

WARNING

TO AVOID DEATH OR SERIOUS INJURY, 
FOLLOW THESE SAFETY INSTRUCTIONS:

Never use the Dräger-Tube unless

– you fully understand and strictly observe these Instructions for Use and the Dräger-

Tube pump's Instructions for Use,

– your employer has trained you how to use the Dräger-Tubes properly, and
– you have passed a test demonstrating to your employer that you can properly use the 

Dräger-Tube!

The Dräger-Tube content is toxic/caustic. Do not swallow. Avoid skin or eye contact.

Use the Dräger-Tubes only for their intended use (see "Intended use" section below)!

Before each measurement, perform a leak test of the pump with an unused Dräger-
Tube; otherwise, you may get a false reading.

After breaking the inner glass ampoule by bending the outer Dräger-Tube, check if 
the heat shrink wrap around the tube 

(2) 

is damaged

If damaged, do not use the 

Dräger-Tube!

Break off 

both

 ends of the Dräger-Tubes before inserting the Dräger-Tube into the 

Dräger-Tube pump because otherwise any measurement will be impossible!

When inserting the Dräger-Tube, the black arrow must point toward the Dräger-Tube 
pump - otherwise, measurement is impossible!

The Dräger-Tube contains a glass ampule, filled with a liquid preparation 

(3)

, which 

you have to break open during the measurement because otherwise any 
measurement will be impossible!

To avoid false readings, never use Dräger tubes after the expiration date printed on 
the packaging or that have been stored in temperatures exceeding the limits printed 
on the packaging!

Use only Dräger-Tube pumps since the use of pumps of other manufacturers may 
cause false readings!

Measuring range:

2 - 15 mg/m

3

Number of strokes (n):

10 (+2)

Standard deviation:

±20 % - 30 %

Color change:

yellow 

 red

Measuring time:

approx. 2.5 min

Correction factor:

F = 1013 hPa (14.692 psi) / actual atmospheric pressure

Temperature:

0 °C - 30 °C/32 °F - 86 °F

Humidity:

< 20 mg/L (20 mg/L correspond to 90 % r. h. at 25 °C/77 °F)

00122930.eps

ampule with liquid preparation,
ampolla con preparado líquido,
ampoule avec une préparation liquide

1

4

1

3

2

yellow indication layer,
nivel de indicación amarillo,
couche indicatrice jaune

filter and separation layer,
filtro y nivel de separación,
couche filtrante et séparatrice

!

es

Para su seguridad

ADVERTENCIA

PARA EVITAR LA MUERTE O LESIONES FÍSICAS GRAVES, 
ES NECESARIO RESPETAR LAS SIGUIENTES INDICACIONES DE 
SEGURIDAD:

Utilizar únicamente los tubos Dräger cuando

– Se haya entendido completamente este manual de instrucciones (tubos Dräger), así 

como el manual de instrucciones de la bomba para tubos Dräger

– Cuando se haya instruido a los usuarios sobre la aplicación técnica del tubo Dräger
– Cuando se haya realizado un test a los usuarios que demuestre que se domina la 

aplicación técnica del tubo Dräger.

El contenido del tubo Dräger es tóxico/corrosivo. No ingerir. Evitar el contacto con la 
piel y los ojos.

Utilizar el tubo Dräger solo según el uso previsto establecido (ver "Uso previsto" a 
continuación).

Realizar antes de cada medición una prueba de estanqueidad de la bomba para 
tubos Dräger con un tubo Dräger todavía sin utilizar para evitar cualquier indicación 
errónea.

Después de romper la ampolla de vidrio interna doblando el tubo Dräger, comprobar 
que la funda protectora del tubo 

(2) 

no presente daños

En caso de daño, no utilizar 

el tubo Dräger.

Romper las 

dos

 puntas del tubo Dräger antes de colocar el tubo Dräger en la bomba 

para tubos Dräger, puesto que de lo contrario no es posible ninguna medición.

Al colocar el tubo Dräger, la flecha negra debe apuntar hacia la bomba para tubos 
Dräger, puesto que de lo contrario no es posible ninguna medición.

El tubo Dräger contiene una ampolla de vidrio rellena de preparado líquido 

(3)

, que 

debe romperse durante la medición, de lo contrario no es posible ninguna medición.

Para evitar indicaciones de error, utilizar solo los tubos Dräger que todavía no hayan 
caducado (ver envase) y que hayan sido almacenados según el rango de 
temperatura indicado (ver envase).

Emplear únicamente bombas para tubos Dräger, puesto que el uso de bombas de 
otro fabricante puede llevar a error.

!

Instructions for Use - 67 28 791
Draeger-Tubes

® 

Cyanide 2/a

© Dräger Safety AG & Co. KGaA · Edition 10 - January 2013 · Subject to alteration · 

90 22 930 

enus_es_fr 

Dräger Safety AG & Co. KGaA · Revalstraße 1 · D-23560 Lübeck · Germany · Tel. +49 451 882 - 0 · Fax. +49 451 882 - 20 80 · www.draeger.com

Instrucciones de uso - 67 28 791

Tubo de control Dräger

®

 Cianuro 2/a

Notice d'utilisation - 67 28 961

Dräger Tube réactif

 ® 

Cyanure 2/a

Отзывы: