
13
4
Maintenance
4.1
Calibration
4.1.1 Calibration Procedure
Calibration of this unit must be performed at regular intervals as detailed in the
sensor data sheet.
DraegerSensor IR:
If the zero signal has drifted significantly, or a calibration fails: calibrate the
DraegerSensor IR first (see manual P/N 9023843), then continue with the
calibration procedure for Polytron FX.
4.1.1.1 Zero Calibration
• Attach the pressure regulator to the nitrogen (N2) or Zero Air calibration gas
cylinder.
DraegerSensor PR and LC:
• Fit the calibration adapter tightly to the end of the sensor.
Draeger Sensor IR:
Maintenance
Calibration
Figure 7: Calibration
Содержание 4543445
Страница 1: ...Polytron FX Operating Manual...
Страница 22: ...21 Certifications...
Страница 23: ...22 Certifications...
Страница 24: ...23 Certifications...
Страница 25: ...24 Certifications sales norrscope com...