Dräger Medizintechnik GmbH
Maintenance Procedure Savina Version 2.0
Page 65
0DLQWHQDQFH3URFHGXUH
5HSODFLQJWKHEORZHUXQLW
• Push the side panel away slightly and carefully withdraw the blower out of the unit towards the
rear.
• Remove the fixing screws
(screws are tighten to 1.2 Nm
±
0.3 Nm).
)LJ
Removing the blower unit.
• Remove the blower block.
When withdrawing the blower pay attention to the cable connections. The cable
connections must also be fed out. The removed cables are those of the Control PCB
and the Motor Commutator PCB.
6
6
6
6
6
6
6
6