Dräger Medical AG & Co. KGaA
Software download Savina Version 3.0
Page 307
6RIWZDUHGRZQORDG
'RZQORDG
'RZQORDG
This description applies with effect from download software 9.1.1.
• Connect Savina and PC (laptop) to mains power supply.
• Connect PC and Savina to RS232 extension and to RS232 adapter.
• Switch on the PC.
The optional ”BIPAP” and ”Autoflow” functions will be lost when updating the software
from SW 1.00 to SW 1.10 or from SW 1.01 to SW 1.10. They must be reactivated by
entering the release number (release code). See also ”
” or ”
SW 1.01 to 1.10 update procedure
” in chapter ”Repair Information/Error
Events”.
7KH6DYLQDLQLWLDOSURJUDPORDGHUFRXOGEHGDPDJHGLQWKHHYHQWRIDSRZHU
IDLOXUH'RQRWUXQWKHGRZQORDGLIWKH6DYLQDDQGRUWKH3&DUHEHLQJSRZHUHGE\
EDWWHU\7KHEDWWHU\SRZHUWRWKH6DYLQDLVQRWVXIILFLHQWIRUWKHGRZQORDG
The software cannot be loaded with a normal RS232 cable. Pins 1 and 9 must be
connected to a 10-k
Ω
resistor to enable the Control PCB to detect the SERVICE-Q
signal. Use only the specified cables and adapters.