F
o
r int
e
rnal us
e only
.
Copyright
res
e
rved.
R
566
4900F
L2.f
m
03.
12.01
Dräger Medizintechnik GmbH
Alarm Messages Savina Version 2.0
$ODUP0HVVDJHV
0RQLWRULQJRIFHQWUDOV\VWHP
Page 263
.H\SDG
&DXVH
1.
The user has pressed the key too many times.
2.
The user keeps the key pressed all the time.
3.
Hardware error (bouncing, short-circuit, open-circuit).
'HWHFWLRQ
1.
The input value "Key XX overused" shows an error.
2.
The input value "Key XX failed" shows an error.
6XSSUHVVLRQ
1.
None.
5HDFWLRQ
1.
Detection 1: Generation of event "Key XX overused" (caution "!! Key XX overused").
2.
Detection 2: Generation of event "Key XX failed" (caution "!! Key XX failed").