Alarm Messages Savina Version 2.0
$ODUP0HVVDJHV
0RQLWRULQJRIVHQVRUV
Page 209
0RQLWRULQJRIVHQVRUV
$LUZD\SUHVVXUHVHQVRUV
&DXVH
1.
A pressure sensor is faulty, e.g. diaphragm is damaged.
2.
Offset voltage of one sensor or both sensors has drifted too much.
3.
The sensors are no longer supplied with voltage.
4.
The calibration valve of one sensor is faulty and, as a result, the sensor is permanently
switched to atmospheric pressure.
5.
A pressure sensor delivers incorrect, but dynamic measured values.