Alarm Messages Savina Version 2.0
$ODUP0HVVDJHV
9HQWLODWLRQPRQLWRULQJ
Page 178
0LQXWHYROXPH
09ORZ
&DXVH
1.
Spontaneous ventilation of the patient is insufficient.
2.
Spontaneous ventilation of the patient is sufficient, but the expiratory flow measured is too
low due to leakages in the expiratory tube.
3.
The flow control unit is faulty and therefore delivers too low an inspiratory flow.
4.
The flow sensor is faulty and therefore measured values delivered are too low.
'HWHFWLRQ
1.
The measured value MV(t) is below the lower MV limit value MV low.
6XSSUHVVLRQ
1.
After a cold start and 2 minutes afterwards.
2.
During standby and 2 minutes afterwards.
3.
When flow monitoring is off and 2 minutes afterwards.