![Dräger Medical Narkomed GS Скачать руководство пользователя страница 53](http://html1.mh-extra.com/html/drager-medical-0/narkomed-gs/narkomed-gs_setup-and-installation-manual_2527076053.webp)
6-1
PMC PROCEDURE
NMGS
6.0
PMC PROCEDURE, NARKOMED GS
The procedures in this section shall be performed in their entirety each time a
component is removed, replaced, calibrated, adjusted and during all scheduled
Periodic Manufacturer’s Certification (PMC) visits. A PMC Checklist form, P/N
S010211 is available from Draeger Medical, Inc. and shall be completed by the
Technical Service Representative each time a PMC is performed. Steps in the
procedure marked with
(
9
)
require a response at the corresponding line on the
checklist form.
Space is also provided on the PMC checklist form to record the results of a vapor
concentration test. Refer to the current Anesthesia Equipment & Monitoring System
Service Information CD-ROM Service Procedures section for vapor concentration
verification procedures.
NOTE:
Test equipment listed below with an asterisk (
*
) requires calibration
at a maximum interval of one year. Verify the dates on test
equipment calibration labels. DO NOT USE any test equipment
having an expired calibration date. Notify your supervisor
immediately if any equipment is found to be out of calibration. In the
space provided at the bottom of the PMC checklist form, record the
Model and ID number of all calibrated test equipment used.
In the space provided at the bottom of the PMC checklist form, record the Model and
ID number of all calibrated test equipment used. Also record the calibration due dates.
Examples are: multimeter, digital pressure meter, Riken gas analyzer, safety analyzer,
volumeter, trace gas analyzer, simulators.
Test Equipment Required:
--
*
Electrical Safety Analyzer (Biotek 501 Pro or equivalent)
--
*
Pressure Gauge with DISS Adapters (P/N 4114807 or equivalent)
--
*
Flowmeter 0-250 ml min. (P/N S000081 or equivalent)
--
*
Volume Meter (P/N 2212300 or equivalent)
--
*
Digital Pressure Manometer (SenSym PDM 200CD or Equivalent)
--
*
Riken Gas Indicator (Model 18H, or 1802D or equivalent)
-- Stop Watch
-- Test Lung (P/N 8401892)
-- AC Receptacle Circuit Tester
Materials Required:
-- Spiromed Lubrication Kit (P/N 2218180)
-- Breathing Bag 3 liter (P/N 9995330 or equivalent)
-- Patient Circuit: Y-piece, elbow, 2x 32” x 22mm hoses
-- Hose 22 mm x 32” (P/N 9995132)
-- Fresh Gas Outlet Volume Test Device (P/N S010158 or equivalent)
Содержание Narkomed GS
Страница 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 10: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 40: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 46: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 50: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 52: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 90: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 91: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...