18. Trouble-Shooting
Dr. A. Kuntze
I
Gutes Wasser mit System
Robert-Bosch-Str. 7a
40668 Meerbusch
+49 (0) 2150-7066-0 / Fax -60
www.kuntze.com
77
18.3. Diagnosis file
In case of problems, it might be useful to provide an external contact person with an
overview of the current situation. To that purpose save the diagnosis file on your SD
card. It contains all current values and settings and the eventlog.
18.4. Info
The Menu SYSTEM=>INFO contains all information on your instrument, the settings,
and the current values, such as:
ID
Here you can find the instrument number, the software versions, and the operation
hours. This information is important for repair, updates, etc.
Eventlog
The eventlog lists up to 100 events with the times of their appearance and
disappearance. Further details are supplied on the following pages.
Settings
This menu lists all settings. In case of trouble, check if all settings are as you had
intended, and check if the instruments behavior is consistent with the settings. If for
example the alarm relay is not activated when the measured value drops below the
lower limit, this might be because you have not activated the lower limit as an alarm
action in the alarm action list.
View
This menu shows all current values of analog and digital inputs and outputs,
including the raw signals of the sensors. The later are invaluable for trouble-shooting
since they are not affected by any user settings such as calibration, temperature
compensation etc. If for example you get a slope error after calibration, even though
the raw signals of the sensor were plausible, then the problem might well be a