7. Temperature measurement
Dr. A. Kuntze
I
Gutes Wasser mit System
Robert-Bosch-Str. 7a
40668 Meerbusch
+49 (0) 2150-7066-0 / Fax -60
www.kuntze.com
35
7.2. Temperature compensation
– ORP measurement
Temperature has a strong influence on the oxidation/reduction potential of most
solutions. But since the ORP potential is often used as such and not as a measure
for a concentration, there is no temperature compensation for ORP measurements.
Therefore the parameter reference
is inaccessible with ORP measurements.
However, temperature measurement is still possible, and the measured temperature
can be displayed in the desktop.
If you have connected a Pt100 or Pt1000, select auto for temperature mode. The
measured values are then shown in the display and are also available via mA or
digital outputs.
If you have no temperature sensor connected, select manual compensation, and
enter the temperature manually. This value will then be displayed on the desktop
and be given out via the various interfaces.