background image

13

12

5.7.1 The Special Function of the Mode Key
5.7.1.1 Inserting Markers

At particular events during the recording 

to hold, 

the sex 

Brands are set. To place a marker (marker) in the graphics of the PDFs, the mode key 
must be held for 2.5 seconds (see marker on the PDF image Fig.C). The successful 
marking is confirmed by a short signal tone. The marking will then be stored with the next 
measurement (note the interval).

5.7.1.2 Removing the MAX-MIN display memory

The logger has a MIN / MAX function for holding the extreme values ​​over a given time frame. Press and 
hold the MODE key for 5 seconds until a small tune is reached. Damit is reset to the measurement period. 
They can, for example, capture the highest / lowest daytime temperatures. Die MIN / MAX Function works 
independently of Log operation.

Please note: 

-

At the start of recording, the MIN / MAX memory is reset to indicate the MIN / MAX / average 
values ​​appropriate for recording.

-

During the recording, when setting the MIN / MAX / average values, a single mark 
will be set.

5.7 Display modes and mode key: EXT, AVG, MIN, MAX

The first line shows the current measurement value of the main measurement 

channels: LOG200: Temperature, LOG210: Humidity, LOG220: Air pressure, 

LOG200 TC / LOG210 TC: External temperature 1, LOG200 E: Internal 

temperature, LOG220 E: Air pressure

The display of the second row depends on the selected mode. There are two display modes:

1. AUTO MODE (

 Preset for LOG200, LOG210, LOG220) - the second line alternates 

1. AUTO MODE (

 Preset for LOG200, LOG210, LOG220) - the second line alternates 

every 3 seconds :: LOG200: 

Minimum (MIN) / Maximum (MAX) / Average (AVG) LOG210: 

Temperature and Dew Point (Td) 

LOG220: 

Humidity, temperature and dew point LOG200 TC: Internal 

temperature and external temperature 2 
LOG210 TC: Internal Temperature, Internal Humidity, Internal Dew Point, External Temperature 2 LOG200 E: 

Current Temperature / Minimum (MIN) / Maximum (MAX) / Average 

(AVG) temperature (external) and minimum (MIN) / maximum (MAX) / average 
(AVG) temperature (internal)

LOG220 E: Current measurement values ​​of external connected sensors / current measurement values 

internal measurement channels humidity, temperature, dew point You recognize the value in 

the physical unit over the measurement value (° C / ° F = temperature, Td + ° C / ° F = dew point,% rH = 

humidity, hPa = pressure, ppm = CO

 2)

humidity, hPa = pressure, ppm = CO

 2)

as well as the additional symbols <no symbol> = current measurement value, MIN = minimum, MAX = 
maximum, AVG = average, EXT = external channel 2, EXT flashing = external channel 1.

The AUTO mode gives you an overview of the measurement values ​​of all channels in a short 
time. By pressing the MODE key (left key) they go from AUTO mode to manual mode:

2. MANUAL Mode (

 Preset for LOG200 TC, LOG210 TC) - the MODE key flashes in the 

2. MANUAL Mode (

 Preset for LOG200 TC, LOG210 TC) - the MODE key flashes in the 

second row through all available measurement channels in the order of current 
measurement value (no symbol), minimum (MIN), maximum (MAX), average (AVG). 
Details:

-

For the main mass channel in the upper row, only MIN, MAX, AVG are displayed.

-

For LOG200 TC and LOG210 TC, MIN, MAX, AVG are indicated with a flashing EXT for 
the external sensor 1 (main channel) and a standing EXT for the external sensor 2.

The manual mode allows the display of an optional measurement channel in addition to the main 
measurement channel, for example, the maximum pressure for the current pressure. Press and hold the 
MODE key until the display AUtO appears to return to AUTO Mode.

Содержание LOG 200

Страница 1: ...210 TC LOG 200 E LOG 220 E PDF data logger for temperature humidity pressure and CO2 PDF data logger for temperature humidity pressure and CO2 PDF Data logs for temperature humidity pressure and CO2 PDF Data logs for temperature humidity pressure and CO2 PDF Registrar of Température Humidity Pressure and CO2 PDF Registrar of Température Humidity Pressure and CO2 5005 0200 5005 0210 5005 0220 5005 ...

Страница 2: ...tton Taste Button Taste Button Stop time Stop time Stop time Stop time Stop time Stop time 27 09 2016 12 43 27 Stop by Stop by Stop by Stop by Stop by Stop by USB connection USB connect USB connection USB connect USB connection USB connect Data sets Records Data sets Records Data sets Records 265 Duration Duration Duration Duration Duration Duration 02 12 00 Min Avg Max Höhe ü dN Altitude ASL Höhe...

Страница 3: ...o clean the instrument do not use an aggressive cleaning agent but only with a dry or damp cloth There is no liquid in the interior of the dish Store the measuring instrument in a dry and clean place Avoid violent effects such as size or pressure For no correct or incomplete measurements and no consequences no gun exists The liability for consequential consequential damages is excluded Do not use ...

Страница 4: ...e device is required FIG A see page 2 FIG A see page 2 1 USB protective cap 2 USB connection 3 LCD display see FIG B 4 LED red green 5 Sensor compartment Position of Sensors 6 Battery compartment 7 Fashion key 8 Start Stop key 9 closable wall mount 10 External connectors only TC 11 External connector only E FIG B see page 2 FIG B see page 2 1 Units in Measurement Value Line 1 2 Measurement value L...

Страница 5: ...oggers to the PC and activate AutoSave default this automatically saves the measurement data 5 2 2 Windows Software for LOG200 Series The LOG200 series data loggers are pre programmed and can be installed without any software It still provides a free software that can be downloaded directly from the following website See 5 2 2 1 LogConnect configuration software 5 2 2 1 LogConnect Configuration So...

Страница 6: ... TC LOG210 TC LOG200 E LOG220 E Open and secure PDF PDF will be overwritten at the next launch NOTICE At the logger start the Extreme Max and Min Value as well as the average NOTICE At the logger start the Extreme Max and Min Value as well as the average value were reset 5 6 USB connection To configure or evaluate the data loggers the data logger must be connected to the USB connection of your PCs...

Страница 7: ...ure and Dew Point Td LOG220 Humidity temperature and dew point LOG200 TC Internal temperature and external temperature 2 LOG210 TC Internal Temperature Internal Humidity Internal Dew Point External Temperature 2 LOG200 E Current Temperature Minimum MIN Maximum MAX Average AVG temperature external and minimum MIN maximum MAX average AVG temperature internal LOG220 E Current measurement values of ex...

Страница 8: ...fter date time query memory interval INT can be changed N No acquires setting Y Yes new settings are stored NOTICE After the battery change please check the clock internal clock and correct it if NOTICE After the battery change please check the clock internal clock and correct it if necessary See 5 2 2 1 Log Connect configuration software or 5 8 1 Set date time after battery change by keyboard The...

Страница 9: ...0 LOG220 E Area of measurement 300 1100hPa Resolution 0 1 hPa Memory LOG200 150 000 data sets temperature and motion LOG210 100 000 data sets temperature relative humidity dew point and motion LOG220 60 000 data sets temperature relative humidity pressure dew point and motion LOG200 TC 60 000 data sets Temperature 2 external thermocouples T1 T2 movement LOG210 TC 60 000 data sets humidity temperat...

Страница 10: ...any liquid into the interior of the device Please store the measuring instrument in a dry and clean place Avoid any force like shocks or pressure to the instrument No responsibility is taken for irregular or incomplete measurement values and their results the liability for subsequent damages is excluded Do not use the device in explosive areas Danger of death Do not use the device in an environmen...

Страница 11: ...ation strictly Use only battery type LS 14250 3 6 volt from manufacturer SAFT or DYNAMIS Lithium Batt LI 110 1 2 AA S respectively only batteries authorized by the manufacturer For operation take out the instrument from the packaging remove the display foil The logger is already preset and ready to start It can be used immediately without any software By pressing any button or moving the instrumen...

Страница 12: ...cording Press button for 2 seconds beeper sounds for 1 second the actual date and time will be displayed for 2 further seconds LED lights green for 2 sconds logging has started LED blinks green every 30 sec Display in Auto Mode Display shows all measurement channel in a 3 second sequence By using the LogConnect software the presets can easily be changed Download free LogConnect software www dostma...

Страница 13: ...DF and save PDF will be overwritten with next log start Note With the next recording the Extrema Max and Min value and the AVG value will be Note With the next recording the Extrema Max and Min value and the AVG value will be reset 5 3 2 Display used memory date and time After pressing the start button while the logger is recording the first line shows MEM and the second line line 2 displays the u...

Страница 14: ...0 Temperature LOG210 Humidity LOG220 Air pressure LOG200 TC LOG210 TC External temperature 1 LOG200 E Internal temperature LOG220 E Internal air pressure The second line display depends on the selected mode There are two display modes 1 AUTO mode default for LOG200 LOG210 LOG220 the second line shows alternating 1 AUTO mode default for LOG200 LOG210 LOG220 the second line shows alternating values ...

Страница 15: ...ly LOG220 LOG220 E Pressure only LOG220 LOG220 E Pressure only LOG220 LOG220 E Measuring range 300 1100hPa Resolution 0 1 hPa Memory LOG200 150 000 measurements Temp Exercise LOG210 100 000 data set Temp Humidity dewpoint and motion LOG220 60 000 data sets Temp Humidity pressure dewpoint and motion LOG200 TC 60 000 measurements Temp 2 external thermocouples T1 T2 motion LOG210 TC 60 000 measuremen...

Страница 16: ...ontained are Cd cadmium Hg mercury Pb lead Table des matières 1 Introduction 32 2 Councils Consignes of Security 32 3 Content 32 4 Distribution a utilization 32 5 Manipulation of the appliance 33 5 1 Description de l appareil 33 5 2 Function of the appliance 34 5 2 1 Installation standard à la livraison 34 5 2 1 1 Mode of Displaying Power Save 35 5 2 2 Logical Windows for LOG200 35 5 2 2 1 Logical...

Страница 17: ... geographical situation and the conditions of the transport or the process are sensitive to the temperature 5 Manipulation of the appliance 5 1 Description de l appareil to humidity and or to pressure The LOG210 TC is an excellent choice for the formation of moississures sur les murs Registering a USB port directly to your PC connects to a PC running on a Windows system tablets or Apple s regulato...

Страница 18: ...e après chaque une heure An appuyant sur n importte quel bouton ou and secouant l registrar les mesures sont reprises toutes les 6 seconds Determination possible pair Bouton Arrest possible couple Connexion USB Alarms arrêtée Retarded the alert 0 s Display of value measurement a march Mode of display Power Save and a march 5 3 Premier allumage debit registration Maintains the bouton appendant pend...

Страница 19: ...stration generation of PDF configuration standard Connect l register to USB interface Le beep sonor est émis pendant 2 seconds environ Une lumière verte clignote jusqu à la génération du PDF which takes just 40 seconds The symbols no sera plus posters dance to the bar The registrar s arrest The beep sonor s fact is switching on the LED light The registrar s present comme un supportable LOG200 LOG2...

Страница 20: ...e internal humidity internal point de rosée internal external temperature 2 LOG200 E température current minimum MIN maximum MAX average AVG of the external temperature LOG220 E Valuation of current measurement of external sensors Value of current measurement of measurement of internal humidity temperature point of view Recognition of value posters in the unit of physical value measurement C F tem...

Страница 21: ...ys The battery of the poster displays a selection of cases in three segments Lorsque l icône de la battery a l ecran clignote a veut dire que la Charge the battery with no surcharge for prendre des mesures 5 7 1 Features of the bouton Mode 5 7 1 1 Rules for markers Pour retention of events by private pendants of registration by way of regulator of markers You place the marker in the PDF graphic as...

Страница 22: ...ter clignote with intermittent pour signals to you material Téléchargez le logiciel gratuit LogConnect www dostmann electronic de home html Download Software voter enregister Software LogConnect_XXX zip XXX choisissez la dernière version 5 8 1 Ruler Date Height of the keyboard for the change of battery regulator Date Heur of the keyboard For a change of battery or rupture and alimentation the devi...

Страница 23: ...remittance of appliances and the fine of the point of recuperation of the electrical and electronic equipment for the purpose of rebuilding so compatible with the protection of the environment 9 Characteristic exclusive to LOG200 Signs CE EN 12830 EN 13485 Storage capacity S and transport T For distribution and distribution C precision class 1 30 70 C Conformance to L En 13486 nous preconisons to ...

Страница 24: ...te for 2 seconds the registration is determined The symbol RUN device Attend queue register so connect to the USB interface attendre Present and secure PDF 1 2 3 4 5 Quick Start with default settings Quick Start with default settings Start Recording Press until beep sounds LED lights green for 2 seconds recording started RUN icon appears Insert logger into USB port wait View and save PDF 1 2 3 4 5...

Страница 25: ...s any errors and misprints reserved Reproduction Stand02 1903CHB DOSTMANN electronic GmbH Technical changes any errors and misprints reserved Reproduction is prohibited in whole or in part Stand02 1903CHB DOSTMANN electronic GmbH Stand02 1903CHB DOSTMANN electronic GmbH DOSTMANN electronic GmbH Messen DOSTMANN electronic GmbH Messen und Steuertechnik Waldenbergweg 3b D 97877 Wertheim Reicholzheim ...

Отзывы: