14 – Calibration

VLM-40 NTSC T

ECHNICAL

I

NSTRUCTION

M

ANUAL

CALIBRATION

NTSC C

ALIBRATION

...

The VLM-40 has been carefully calibrated for NTSC operation
at the factory. However, the procedure is listed here for your
future use.

NOTE: The VLM-40 is only as accurate as the waveform generator
you are using. Check sync and raster amplitudes before beginning
calibration.

1. With the power off, remove the bottom four screws and lift

off the enclosure to expose the circuit board, as shown in
Fig. 17 (on the next page).

2. Connect a waveform generator to the VLM-40 video input

and turn on the termination switch, SW1. Set the waveform
generator to split-field bar pattern (100% white flag). 

3. Apply power to the VLM-40. Adjust the rear panel input

level control VR 13 so that the 100 IRE peak dot LED is lit.
Verify that the -40 IRE sync amplitude LED is lit.

4. Select the full-field white raster waveform (100% white) on

the generator. On the main board, adjust the average
calibration control VR 36 so that the 100 IRE average bar
LED is lit.

5. Select the stairstep waveform on the generator. Verify that

the set-up LED on the VLM-40 is now lit red, indicating a
set-up violation.

6. Disconnect the waveform generator from the VLM-40.

Verify that both the sync and set-up error LEDs are red, and
that the main display is blanked.

7. Turn off power and reassemble the unit.

Содержание VLM-40 NTSC

Страница 1: ...COMPOSITE VIDEO LUMINANCE METER MODEL VLM 40 48 40 0 20 40 60 80 100 120 1990 SYNC LUMINANCE IRE UNITS 32 SYNC SETUP COMPOSITE VIDEO LUMINANCE METER MODEL VLM 40 NTSC TECHNICAL INSTRUCTION MANUAL...

Страница 2: ...ret data on the single meter scale The VLM 40 is insensitive to chroma information VITS vertical interval test signals and it responds to luminance signals as short as 200 ns in length INPUT IMPEDANCE...

Страница 3: ...ncludes sync and setup LEDs that detect the status of these corresponding video levels Under normal operating conditions the sync LED will be green whenever the video source has a proper sync signal I...

Страница 4: ...be installed in an optional panel mount e g DORROUGH Rack Mount Kit 40 S for a single meter or 40 D for dual meters or operated as a stand alone package The remainder of this Technical Instruction Man...

Страница 5: ...VLM 40 refer to Fig 3 for mounting dimensions Choose a ventilated area that provides a minimum depth of 10 for proper cable routing Consider a location that not only will offer easy observation of the...

Страница 6: ...eo BNC connector on the back panel Select Term or Hi Z to terminate the connection 4 Feed a standard EIA split field color bar signal with standard setup at 140 IRE 1 V p p to the VLM 40 and verify th...

Страница 7: ...ture tube as shown in Fig 5 The synchronization signals embedded in the television signal precisely instruct the electron beam to move across and down the picture tube to build a complete picture usin...

Страница 8: ...ve picture lines occurring after 21 lines of vertical blanking For PAL SECAM measurements contact the the factory for information on the VLM 40 PAL SECAM meter 1 Using a standard television test signa...

Страница 9: ...60 40 20 20 40 0 setup 7 5 COMPOSITE VIDEO LUMINANCE METER MODEL VLM 40 48 40 0 20 40 60 80 100 120 1990 SYNC LUMINANCE IRE UNITS 32 SYNC SETUP COMPOSITE VIDEO LUMINANCE METER MODEL VLM 40 48 40 0 20...

Страница 10: ...k luminance on a waveform monitor as shown in Fig 10 NOTE The VLM 40 is very sensitive to measuring noise in the luminance signal and will respond to short impulses that may be undetected on a standar...

Страница 11: ...in Fig 12 The thresholds are Setup Level Setup LED Above 7 5 IRE Green 7 5 to 4 0 IRE Green Below 4 0 IRE Red COMPOSITE VIDEO LUMINANCE METER MODEL VLM 40 48 40 0 20 40 60 80 100 120 1990 SYNC LUMINA...

Страница 12: ...E IRE UNITS 32 SYNC SETUP IRE Units 100 80 60 40 20 20 40 0 setup 7 5 Fig 12 On the VLM 40 the sync LED flashes red each time a sync pulse is missing You may not easily detect this fault using a wavef...

Страница 13: ...vel is less than 4 IRE IRE Units 100 80 60 40 20 20 40 0 setup 7 5 COMPOSITE VIDEO LUMINANCE METER MODEL VLM 40 48 40 0 20 40 60 80 100 120 1990 SYNC LUMINANCE IRE UNITS 32 SYNC SETUP Fig 15 On the VL...

Страница 14: ...VLM 40 the average luminance levels are below 40 IRE and the peak LED indicates peak luminance levels are at 60 IRE The waveform monitor shows peak luminance levels just above 60 IRE but calculating a...

Страница 15: ...on the termination switch SW1 Set the waveform generator to split field bar pattern 100 white flag 3 Apply power to the VLM 40 Adjust the rear panel input level control VR 13 so that the 100 IRE peak...

Страница 16: ...n 15 Fig 17 This simplified circuit board diagram shows key components used to calibrate the VLM 40 for NTSC operation VLM 40 NTSC TECHNICAL INSTRUCTION MANUAL CALIBRATION TRANSFORMER J2 J1 VR36 U7 U3...

Отзывы: