40
Fig.30. Storage mode indication
By pressing button
ОК
RealTime_PCR
software is switched to
Archive viewer
mode
(
Data analysis
tab).
If option
“Turn off device after the run”
was activated, then upon completion of the RUN
the instrument switched to the sleep mode. To exit this mode, launch RealTime_PCR
software and again connect to the instrument in the appeared window
List of devices
.
Note!
To read results of the executed RUN after exit sleep mode proceed to menu
Preferences/Last run in device
option.
The analysis of optical measurements results are described in details in the second part of
operation manual
Software guidance
.
5. Maintenance
5.1 General Guidelines
The maintenance of the instrument is necessary to keep it at optimum working condition,
which minimises the risk of having unscheduled downtime and providing for its maximum
operational life.
The instrument belongs to IVD medical equipment. Maintenance should be carried out only
by competent personnel, who have studied the present operation manual.
The instrument is required for minimal routine maintenance during normal laboratory
operation.
Caution, risk of electrical shock!
Replacement of fuses should be done on de-
energized instrument only. Fuses type is T10AH250V with the following parameters - 10A,
250V (slow with high breaking capacity) and 5x20 mm size should be used in the case of
their replacement. De-energizing of the instrument is achieved by removing the power cord
connector from it or from the power outlet and disconnection of USB cable as well.