Power Switch Alarm Reset

The power switch simply switches the mains power to the pump on and off.

When the pump detects an alarm LED condition, this can be cancelled by switching the pump off and then back on again.

CPR

Please ensure that the CPR tag is always positioned fi rmly into the mattress (Closed), prior to infl ating the mattress. 

The mattress will NOT infl ate should the CPR tag not be closed. The CPR tag is only to be used in the event of a clinical 

emergency for priority use. However the function can also be used to defl ate air rapidly from the mattress in readiness 

for transport. 

Settings 

When the system is initially switched on using the power switch as above,simply select the valve from soft to fi rm as 

clinically required. Patients who are up to the maximum user weight (24 stone) should be placed on a much fi rmer 

setting than a frail or light weigh patient Please be patient and wait until the system fully infl ates from initial installation. 

This may take up to 15 minutes from the initial set up.

Normal Pressure 

When the system is operating at a normal functioning pressure, then the LED should always be shown as 

highlighted left.

Low Pressure 

If the mattress is operating at too low a pressure setting, there is a mattress “leak” or indeed if the CPR has not been 

closed fully, then a low pressure light will be indicated as shown.

Quick Release Connector

The connectors (male & female coupling) as indicated left can also be used to create a “static mode”. This is achieved by 

pressing home the male & female section together until locked tight To release the connectors, simply press the metal 

release catch.

Filter

The fi lter is situated at the rear of the pump. This can be removed and cleaned whilst on site to improve infection 

control measures. Please DO NOT insert the fi lter back into the unit when still wet Electricity and water is a dangerous 

combination.

Dyna-Form Air System

D I R E C T H E A L T H C A R E S E R V I C E S . C O . U K

Q U I C K   R E F E R E N C E   G U I D E

This is a quick reference guide for the 

Dyna-Form Air System

 

(refer to user manual for additional information). 

Product Code MAT/AIR/200/85/10

Controls