10.6 Cleaning and Disinfection of Dürr spittoon valve on wet suction system
(optional)
If DIPLOMAT unit is equipped with Dürr spittoon valve,
the following are required for disinfection/cleaning:
Material-compatible, non-
foaming disinfection/cleaning agents that have been approved by Dürr Dental,
e.g. Orotol plus or Orotol Ultra.
The procedure
1.
Switch on the rinsing for the spittoon.
2.
Keep pressing the cleaning button on the yellow switch control panel,
until rinsing of the spittoon is finished.
3.
Pour disinfection solution into the spittoon and at the same time press
the yellow cleaning button on the switch control panel, until the disinfection
solution has been aspirated.
Monthly maintenance:
Press the cleaning button to empty the collection vessel.
Clean the yellow coarse filter or replace it if required.
The yellow coarse filter prevents larger dental particles from reaching the
suction pipe.
NOTICE:
Device malfunctions or damage due to use of incorrect media. Guarantee claims may become invalid as a
result.
•
Do not use any foaming agents, e.g. household cleaning agents or instrument disinfection agents.
•
Do not use abrasive cleaners.
•
Do not use agents containing chlorine.
•
Do not use any solvents like acetone.
Cleaning Button
coarse filter