INSTRUCTIONS FOR USE
DA 170, DA 130
UM_EN_DA170_DA130_2018-09_ver2.6
10
/
37
Air
Oil-free, clean and dry air, with a minimum flow of 55 l/min and a pressure of 0,45 to 0,8 MPa.
Suction
(for cuspidor block equipped with large and small aspirators
)
Static vacuum index must be measured at the spot and must be within the range of min. 0,005 MPa (50 mbar) to max.
0,02 MPa (200 mbar). If the static vacuum index is higher than 0,02 MPa, then suction CONTROL valve should be
connected to the suction branch in order to limit the max. vacuum to 0,02 MPa. This regulating valve is not a part of the
kit. The suction unit must produce the flowrate of at least 450 Nl/min. measured at the spot.
Waste
The waste /drain/ branch must have continuous slope of min. 1% and min. flowrate of 10l/min. and must have no sharp
bends and sections that might cause backflow. Do not use the same waste branch in conjunction with another dental
unit or a basin. It is allowed to use polypropylene or cured polyethylene tubes.
Note
If the local regulations require an installation of an amalgam catcher, then the cuspidor block which is not
equipped with the amalgam separator, must be connected to an external amalgam catcher. Amalgam catcher
should be installed according to the manufacturer's instructions for the product.
Recommended mains fuse rating
Recommended fuse rating for the supply main is 16A. (If using a circuit breaker, use circuit breaker "C" type). No other
equipment should be connected to the supply main! Max. Electrical power input of dental unit is 1500 VA. The supply
main must conform to prevailing local codes.
Recommendation
The manufacturer recommends installation of an instantaneous residual-current device with 30mA sensitivity, if
installation of such does not contradict local regulations.
If all the conditions fit the pre-installation requirements, the dental unit can be installed and connected to the utilities.
5.3 Floor surface
The floor must have at least 100 mm thick concrete foundation. The floor slope should not exceed 1%. Antistatic floor is
recommended.
5.4 Environment
Ambient temperature range
from +10 °C
to +40 °C
Relative humidity range
from 30 %
to 75 %
Atmospheric pressure range
from 700 hPa
to 1060 hPa
1 - Input water from an external water supply
2 - Mount point of inlet water
3 - Mount point for taking the samples of the input water
4 - Filter of water particles
5 - Manual shut-off valve