Operating Instructions UVD 170S/340S
14
Problem
Probable cause
Remedial action
Strong baseline drift
Insufficient warm-up time
Allow min. 30 min. warm-
up
Column contaminated
Rinse or replace column
System not equilibrated
Continue to rinse system
until equilibrated
Unstable environment
Ensure constant temperature
and humidity
Solvent contaminated
Change solvent
Flow cell contaminated
Clean flow cell (
→
sec.
Lamp defective
Replace lamp
Solvent reservoir is placed too
low
Place reservoir on the same
level as flow cell
High noise level
Solvent contaminated
Replace solvent
Lamp aged
Replace lamp
Gas bubbles in system
Prime system
Pressure pulsation from pump
Check and prime pump
Solvent reservoir is placed on
the floor
Place reservoir on the same
level as flow cell
Wrong wavelength
Select suitable wavelength
Connecting cable between
UVD 170S/340S and PC not
correctly installed
Connectors on the detector
and the PC must be inserted
evenly and locked into place
with latching devices on both
sides.
Peak broadening
Capillaries too long or has
excessive internal diameter
Shorten or replace for
narrower capillaries
Column
overloaded,
contaminated or aged
Rinse or replace column
Solvent aging
Replace with fresh solvent
Analysis is not
reproducible
Sample is unstable
Use fresh sample or alter
conditions
Irreproducible gradient
Change gradient program /
check pump
Unstable environment
Ensure constant temperature
and humidity
Column
overloaded,
contaminated or aged
Rinse or replace column
CHROMELEON
TM
Error Messages
A list of CHROMELEON
TM
error messages is included in the CHROMELEON
TM
Installation
program on the tab dialog box "Error Levels". See also section 4.3.3.