20. Reset the lamp lifetime as follows:
a. Press F8 to display the Commands dialog box.
b. Select
UV
(to display UV-related commands).
c. Select
UVLampAge
.
d. Reset the lamp age to zero.
e. Click the
Execute
button.
NOTE It is normal for the baseline to drift for several hours
after a new deuterium lamp is installed.
21. Run the wavelength calibration and verification for the new lamp from the