Changing Ionization Modes
Switching from APCI to ESI ________________________________________________________________
Switching from APCI to ESI
The starting point for this procedure is the source setup for APCI operation,
with the LC and gas flows off, and the probe cooled.
WARNING.
Allow the source block and probe heater assembly to cool
before changing ionization modes.
1.
2.
Unscrew and remove the PEEK finger-tight fitting from the APCI probe
assembly (P/N FM102587).
Change the APCI blank plug (P/N FM101437) with the APCI corona
pin (P/N FM101433), located in the holder in the source enclosure door.
Figure 2-7. APCI blank plug and APCI corona pin
3. Turn the locking plate clockwise to the open position, and remove the
APCI probe assembly (P/N FM102587).
2-4
___________________________ MSQ Hardware Manual ____________________________