background image

 

IonPac TCC-LP1, TCC-ULP1, and TCC-XLP1 
Product Manual 

Page 10 of 17 

 

 

 

 

 

 

Doc. No. 034973-06 

©2005 DIONEX 

September 2005 

 

3.3.

 

Concentrator Capacity 

 

3.3.1.1.

 

Capacity Considerations for Concentrators  

As in all ion exchange systems, the resin has a finite capacity. It can strip a given amount of ions from water. When the 
capacity of the concentrator is exceeded, the stripping will not be quantitative. This condition is referred to as column 
overload.  
 
When estimating the capacity of a concentrator, one must remember that the column is used in a dynamic state where the 
liquid containing the analytes is flowing over the resin at a finite rate. This reduces the capacity somewhat since the analyte 
ions have less time to interact with the resin surface.  

 

Low concentrator column capacity creates the following practical implications:  

A.

 

Trace analysis of an analyte is difficult in the presence of µg/L concentrations of species which exhibit higher or 
similar affinities for the resin. If the dynamic column capacity is exceeded, high affinity ions will displace the 
analytes on the ion exchange sites and result in their elution to waste during the loading process. 

 

 

B.

 

Conversely, qualitative analysis of ions with higher affinities for the resin in the presence of high concentrations of 
ions with low affinities is possible. Again, the key to successful analysis is that the ionic content of the high affinity 
ion to be quantitated may not exceed the effective column capacity. 
 

C.

 

Do not dilute samples to be concentrated in eluent because the eluent ions elute the ions of interest.  
 

D.

 

A plot of response versus concentration should be generated as in Figure 5, “Linearity of Lithium Concentrator 
Injection,” for the determination of the maximum amount of sample or standard that can be quantitatively loaded. In 
Figure 4, “Linearity of Lithium Concentrator Injection,” the break in the curve where linearity starts to change is at a 
concentration volume of 1 mL of 250 µg/L lithium. For practical purposes the amount concentrated for a series of 
samples should be 75% of this value. This will ensure that there is a safety margin built into the concentration 
process in case a sample in a series of concentration experiments has a slightly higher ionic concentration.  

 

 

 

FIGURE 5 

Linearity of Lithium Concentrator Injection 

 

Содержание IonPac TCC Series

Страница 1: ...IonPac TCC...

Страница 2: ...ptember 2005 PRODUCT MANUAL FOR TRACE CATION CONCENTRATOR TCC TCC LP1 Column Low Pressure 4 x 35 mm P N 063079 TCC ULP1 Column Ultra Low Pressure 5 x 23 mm P N 063783 TCC XLP1 Column Extremely Low Pre...

Страница 3: ...acity 10 4 EXAMPLE APPLICATIONS 13 4 1 Concentration in the Manual Mode 13 4 2 Concentration in the Manual Mode 14 5 TROUBLESHOOTING GUIDE 15 5 1 High Backpressure from a Contaminated Inlet Bed Suppor...

Страница 4: ...throughout this document to emphasize important points The symbols are shown below along with the purpose of the information SAFETY Safety information can help prevent bodily harm WARNING Warning inf...

Страница 5: ...CC ULP1 is less than 45 psi 0 31 MPa at 1 mL min and the TCC XLP1 is less than 30 psi 0 21 MPa at 1 mL min The large resin particle size 20 m in the TCC LP1 TCC ULP1 and TCC XLP1 makes it possible to...

Страница 6: ...ional Group Hydro phobicity TCC LP1 4 mm 20 55 260 eq col Carboxylic Acid Very low TCC ULP1 5 mm 20 55 260 eq col Carboxylic Acid Very low TCC XLP1 6 mm 20 55 260 eq col Carboxylic Acid Very low TABLE...

Страница 7: ...Column Description The IonPac Trace Cation Concentrator Low Pressure TCC LP1 4x35 mm and IonPac Trace Cation Concentrator Ultra Low Pressure TCC ULP1 5x23 mm columns consist of the following component...

Страница 8: ...ugh to ensure good quantification WARNING The flow direction during the concentration step is critical In order to ensure optimal system performance it is recommended that concentration always be perf...

Страница 9: ...IONEX September 2005 Figure 3 shows the configuration for sample loading using a Rheodyne valve FIGURE 3 Configuration for Determining Trace Levels of Cations Figure 4 shows the configuration for samp...

Страница 10: ...o 8 days Recommended storage vessels are Corning tissue culture flasks The following procedure should be used for storage of g L level samples A Rinse the polystyrene container and cap twice with deio...

Страница 11: ...he analytes on the ion exchange sites and result in their elution to waste during the loading process B Conversely qualitative analysis of ions with higher affinities for the resin in the presence of...

Страница 12: ...e rate at which the sample is loaded C The pH of the sample D The ionic strength of the sample E The amount and capacity of resin in the column The breakthrough volume is determined as follows A Prepa...

Страница 13: ...2005 DIONEX September 2005 FIGURE 6 Linearity Determinations for Concentrator Injection Time min that the simulated sample is pumped through the TCC LP1 TCC ULP1 or TCC XLP1 after the single ion injec...

Страница 14: ...thanesulfonic acid Eluent Flow Rate 1 0 mL min Injection Volume 1 0 mL Temperature 35 C SRS Suppressor Cation Self Regenerating Suppressor CSRS ULTRA AutoSuppression Recycle Mode MMS Regenerant 100 mN...

Страница 15: ...n Eluent 10 mM Methanesulfonic acid Eluent Flow Rate 1 0 mL min Injection Volume 1 0 mL SRS Suppressor Cation Self Regenerating Suppressor CSRS ULTRA AutoSuppression Recycle Mode or AES Suppressor Cat...

Страница 16: ...back on to the column Follow the instructions below to change the bed support assembly using one of the two spare bed support assemblies included in the ship kit provided with the column A Disconnect...

Страница 17: ...D Be sure that the eluents are made from chemicals with the recommended purity see Section 3 Operation E Be sure that the deionized water used to prepare the reagents has a specific resistance of 18...

Страница 18: ...he gradient pump Direct the effluent from the TCC LP1 TCC ULP1 or TCC XLP1 directly to a waste container C Set the flow rate to 1 mL min D Pump the 1 M HCl 0 1 M KCl solution through the column for 15...

Отзывы: