ICS-90 Installation Instructions
20
Doc. 031856-03 12/06
2.5.3
Verifying Communication with the ICS-90
To verify that Chromeleon or Chromeleon Xpress is communicating with
the ICS-90, click the
Disconnect
and then the
Connect
button on the
Control panel while observing the ICS-90 front panel LEDs.
If the ICS-90 and the chromatography software are communicating
correctly, the
Power
and
Ready
LEDs are on when the ICS-90 is
connected to Chromeleon or Chromeleon Xpress, and only the
Power
LED is on when it is disconnected.