9 • DC Service
Doc. 065031-04 1/08
309
3. On the Wellness panel under
External Conductivity Cell
Calibration
, click
Instructions
for a summary of the procedure
(described in detail in this section).
4. Disconnect the pump output line from the injection valve.
5. Disconnect the line from the suppressor
ELUENT OUT
port to the cell
inlet and connect the pump output line directly to the cell inlet.
6. Verify that there is a minimum of 7 MPa (1000 psi) of backpressure.
7. Set the cell heater to 35 °C
.
Allow the cell to reach this temperature
and then wait an additional 5 minutes to let it stabilize.
8. Begin pumping 1.0 mM KCl through the cell at 1.0 mL/min.
9. Wait until the total conductivity reading stabilizes (in about
15 minutes). Open the Wellness panel again and click
Calibrate
.
After calibration, the conductivity reading should be 147.00 ± 2
μ
S. If
this is not the case, contact Dionex for help.
10. To record the new calibration value in the Audit Trail, click the
Log
button for the cell that was calibrated.
11. Flush the KCl solution from the system by pumping ASTM Type I
(18 megohm-cm) filtered and deionized water through the cell. When
the conductivity drops to less than 1
μ
S/cm, stop the pump.
12. Reconnect the pump to the injection valve and reconnect the line
from the suppressor to the cell inlet.
9.23.2 Replacing a Suppressor
Refer to the suppressor manual for guidance about when to replace a
suppressor. Suppressor manuals are included on the Dionex Reference
Library CD-ROM (P/N 053891).
1. Refer to the suppressor quick start guide (shipped with the
suppressor) for instructions on preparing the suppressor before initial
use.
2. Press the
POWER
button on the front of the DP/SP to turn off the
pump.
3. Press the
POWER
button on the front of the DC to turn off the DC.
Содержание ICS-3000
Страница 1: ...ICS 3000 Ion Chromatography System Operator s Manual Document No 065031 Revision 04 January 2008...
Страница 32: ...ICS 3000 Ion Chromatography System 16 Doc 065031 04 1 08...
Страница 93: ...2 DC Description Doc 065031 04 1 08 77 Figure 2 42 DC Control Panel...
Страница 99: ...2 DC Description Doc 065031 04 1 08 83 Figure 2 46 Conductivity Detector Control Panel...
Страница 106: ...ICS 3000 Ion Chromatography System 90 Doc 065031 04 1 08...
Страница 138: ...ICS 3000 Ion Chromatography System 122 Doc 065031 04 1 08...
Страница 166: ...ICS 3000 Ion Chromatography System 150 Doc 065031 04 1 08...
Страница 174: ...ICS 3000 Ion Chromatography System 158 Doc 065031 04 1 08...
Страница 180: ...ICS 3000 Ion Chromatography System 164 Doc 065031 04 1 08...
Страница 218: ...ICS 3000 Ion Chromatography System 202 Doc 065031 04 1 08...
Страница 228: ...ICS 3000 Ion Chromatography System 212 Doc 065031 04 1 08...
Страница 272: ...ICS 3000 Ion Chromatography System 256 Doc 065031 04 1 08...
Страница 296: ...ICS 3000 Ion Chromatography System 280 Doc 065031 04 1 08...
Страница 354: ...ICS 3000 Ion Chromatography System 338 Doc 065031 04 1 08...
Страница 362: ...ICS 3000 Ion Chromatography System 346 Doc 065031 04 1 08...
Страница 368: ...ICS 3000 Ion Chromatography System 352 Doc 065031 04 1 08...
Страница 372: ...ICS 3000 Ion Chromatography System 356 Doc 065031 04 1 08...
Страница 374: ...ICS 3000 Ion Chromatography System 358 Doc 065031 04 1 08...