8 • EG Troubleshooting
Doc. 065031-04 1/08
205
This error occurs if the flow rate is set to a value the EG does not support. The
DP/SP flow rate range is 0.001 to 10.0 mL/min; however, when an EG is
installed, the allowed range is 0.01 to 3.00 mL/min. The recommended
operating range is 0.4 to 2.00 mL/min.
To troubleshoot:
Set the flow rate to a value within the allowed range.
This error occurs if the selected concentration is too high for the current flow
rate.
To troubleshoot:
Set the flow rate to a value within the allowed range. The allowable eluent
concentration for a particular application depends on several factors: the flow
rate, suppressor type, EluGen cartridge type, and cartridge configuration. For
details, refer to
This error occurs when the current applied to the EluGen cartridge exceeds
the maximum current allowed. (The EluGen current is automatically turned
off to prevent damage to the cartridge.) This error may also occur if the liquid
flow to the cartridge is interrupted.
EG1 invalid flow.
-or-
EG2 invalid flow.
EG1 invalid flow rate-concentration.
-or-
EG2 invalid flow rate-concentration.
EG1 over current.
-or-
EG2 over current.
Содержание ICS-3000
Страница 1: ...ICS 3000 Ion Chromatography System Operator s Manual Document No 065031 Revision 04 January 2008...
Страница 32: ...ICS 3000 Ion Chromatography System 16 Doc 065031 04 1 08...
Страница 93: ...2 DC Description Doc 065031 04 1 08 77 Figure 2 42 DC Control Panel...
Страница 99: ...2 DC Description Doc 065031 04 1 08 83 Figure 2 46 Conductivity Detector Control Panel...
Страница 106: ...ICS 3000 Ion Chromatography System 90 Doc 065031 04 1 08...
Страница 138: ...ICS 3000 Ion Chromatography System 122 Doc 065031 04 1 08...
Страница 166: ...ICS 3000 Ion Chromatography System 150 Doc 065031 04 1 08...
Страница 174: ...ICS 3000 Ion Chromatography System 158 Doc 065031 04 1 08...
Страница 180: ...ICS 3000 Ion Chromatography System 164 Doc 065031 04 1 08...
Страница 218: ...ICS 3000 Ion Chromatography System 202 Doc 065031 04 1 08...
Страница 228: ...ICS 3000 Ion Chromatography System 212 Doc 065031 04 1 08...
Страница 272: ...ICS 3000 Ion Chromatography System 256 Doc 065031 04 1 08...
Страница 296: ...ICS 3000 Ion Chromatography System 280 Doc 065031 04 1 08...
Страница 354: ...ICS 3000 Ion Chromatography System 338 Doc 065031 04 1 08...
Страница 362: ...ICS 3000 Ion Chromatography System 346 Doc 065031 04 1 08...
Страница 368: ...ICS 3000 Ion Chromatography System 352 Doc 065031 04 1 08...
Страница 372: ...ICS 3000 Ion Chromatography System 356 Doc 065031 04 1 08...
Страница 374: ...ICS 3000 Ion Chromatography System 358 Doc 065031 04 1 08...