8 • DP/SP Troubleshooting
Doc. 065031-04 1/08
195
To troubleshoot:
1. Check all tubing connections for signs of blockage and replace it as
needed. Be careful not to overtighten fittings.
2. As columns age, their backpressure increases. It may be necessary to
compensate for this by increasing the high pressure limit. If the column is
the source of the high backpressure, clean the column. (Refer to the
column manual provided on the Dionex Reference Library CD-ROM
(P/N 053891) for instructions.) If this does not eliminate the problem,
replace the column.
3. Observe a run to see whether the high pressure limit is triggered when
injection occurs; if it is, the injection valve may be the source of the
blockage (i.e., the rotor seal or stator may need to be replaced). Contact
Dionex for assistance.
This error occurs when the cap on the seal wash reservoir is equipped with a
drop sensor and the number of drops of wash solution detected by the sensor
while the peristaltic pump is idle (55 minutes per hour) exceeds the leak
detection threshold.
To troubleshoot:
1. Inspect the piston seals for leakage. Replace the piston seals and O-rings
2. Increase the value for the rear-seal leak threshold:
a. In Chromeleon or Chromeleon Xpress, select
Command
on the
Control
menu or press the
F8
key.
b. In the Commands dialog box, select the pump name.
c. Select the
RearSealLeakCounter
command.
d. Enter a new value and click
Execute
.
The rear-seal leak count is x.x (counted drops) and has
exceeded the limit of y.y (leak detection threshold).
Содержание ICS-3000
Страница 1: ...ICS 3000 Ion Chromatography System Operator s Manual Document No 065031 Revision 04 January 2008...
Страница 32: ...ICS 3000 Ion Chromatography System 16 Doc 065031 04 1 08...
Страница 93: ...2 DC Description Doc 065031 04 1 08 77 Figure 2 42 DC Control Panel...
Страница 99: ...2 DC Description Doc 065031 04 1 08 83 Figure 2 46 Conductivity Detector Control Panel...
Страница 106: ...ICS 3000 Ion Chromatography System 90 Doc 065031 04 1 08...
Страница 138: ...ICS 3000 Ion Chromatography System 122 Doc 065031 04 1 08...
Страница 166: ...ICS 3000 Ion Chromatography System 150 Doc 065031 04 1 08...
Страница 174: ...ICS 3000 Ion Chromatography System 158 Doc 065031 04 1 08...
Страница 180: ...ICS 3000 Ion Chromatography System 164 Doc 065031 04 1 08...
Страница 218: ...ICS 3000 Ion Chromatography System 202 Doc 065031 04 1 08...
Страница 228: ...ICS 3000 Ion Chromatography System 212 Doc 065031 04 1 08...
Страница 272: ...ICS 3000 Ion Chromatography System 256 Doc 065031 04 1 08...
Страница 296: ...ICS 3000 Ion Chromatography System 280 Doc 065031 04 1 08...
Страница 354: ...ICS 3000 Ion Chromatography System 338 Doc 065031 04 1 08...
Страница 362: ...ICS 3000 Ion Chromatography System 346 Doc 065031 04 1 08...
Страница 368: ...ICS 3000 Ion Chromatography System 352 Doc 065031 04 1 08...
Страница 372: ...ICS 3000 Ion Chromatography System 356 Doc 065031 04 1 08...
Страница 374: ...ICS 3000 Ion Chromatography System 358 Doc 065031 04 1 08...