ASE 150 Operator’s Manual
78
Doc. 065207-02 9/08
5.5.7
Reinstalling the Pump
1. Reinstall the pump on the component panel.
2. Reconnect the red gas line to the left press fitting on top of the pump.
Reconnect the transparent blue gas line to the right press fitting.
3. Reconnect the solvent lines to the inlet and outlet check valve
housings.
4. Reconnect the cable to the pump PC board.
5.5.8
Completing the Procedure
1. Follow the instructions on
to reinstall the right-side panel.
2. Reconnect the gas source at the ASE 150 rear panel and turn on the
gas supply.
3. Turn on the ASE 150 main power switch.
4. Rinse the system (see
) and check that the pump flow is
normal.