5 • Service
Doc. 065207-02 9/08
71
6. Refer to
and the following steps to reassemble the sample cell.
a. Place the new or cleaned frit into the bottom of the end cap.
b. Press a new PEEK seal (P/N 061687, pkg. 50) into the bottom of the cap
insert.
c. Align the pins in the cap insert with the grooves in the end cap and then
place the insert, with the seal facing down, into the end cap.
7. With the cap assembly upright on the workbench, install the snap ring:
a. Insert the snap ring tool into the holes on the ring.
b. Squeeze the tool handles to bring the ends of the ring together.
c. Insert the ring into the cap. Using your fingers, push the ring under the lip
of the end cap. After verifying that the entire ring is under the lip, release
the tension on the tool and remove the tool from the ring.
d. Screw the cap back onto the cell body and hand-tighten.
Figure 5-3. Sample Cell Assembly
Cell Body
Snap Ring
Cap Insert
Seal
End Cap
O-Ring
Cell End Cap Assembly
Frit