ASE 150 Operator’s Manual
50
Doc. 065207-02 9/08
To edit a custom method (for example, method 1):
1. Press
MENU
to display the
MENU
screen.
PURGE TIME
Amount of time the cell is purged with
nitrogen. Dionex recommends the following
settings:
•
40 to 80 seconds for a 1-mL, 5-mL,
10-mL, or 22-mL cell
•
70 to 110 seconds for a 34-mL cell
•
160 to 200 seconds for a 66-mL cell
•
250 to 290 seconds for a 100-mL cell
20 to 900 sec
(default = 100)
STATIC CYCLE
Number of times the static heating and rinsing
steps are performed. When more than one
cycle is specified, the rinse volume is divided
among the cycles.
1 to 5
(default = 1)
CELL TYPE
Specifies the type of cell used.
SST
(stainless
steel) (default)
Zr
(zirconium)
Figure 3-17. Menu Screen
Parameter
Function
Value Range
Table 3-1. Method Parameters (Continued)
STATUS
SETUP
METHOD EDITOR
DIAGNOSTICS
Содержание ASE 150
Страница 1: ...ASE 150 Accelerated Solvent Extractor Operator s Manual Document No 065207 Revision 02 September 2008...
Страница 32: ...ASE 150 Operator s Manual 24 Doc 065207 02 9 08...
Страница 64: ...ASE 150 Operator s Manual 56 Doc 065207 02 9 08...
Страница 76: ...ASE 150 Operator s Manual 68 Doc 065207 02 9 08...
Страница 114: ...ASE 150 Operator s Manual 106 Doc 065207 02 9 08...
Страница 132: ...ASE 150 Operator s Manual 124 Doc 065207 02 9 08...