3 • Operation and Maintenance
Doc. 065207-02 9/08
39
3. Place a new O-ring over the opening in the end of the cell end cap.
Press the O-ring into place using the O-ring insertion tool (see
3.1.6
Installing the Collection Vessel
During the extraction process, sensors determine if a collection vessel is
present and, if so, whether it is full. If you want to attach a label to a
collection vessel, or write an identification name or number on it, be sure
the location does not block the sensors.
shows the acceptable
locations for this information.
Figure 3-9. Acceptable Locations for Collection Vessel Labels
Before each run, carefully inspect all collection vessels for chips,
scratches, or cracks. If a collection vessel shows any sign of damage,
do not use it.
Use each nongraduated collection vessel once only. Graduated
collection vessels may be used multiple times, provided they are
undamaged.
La présence de fêlure ou d’égratignures sur les flacons de collecte
doit être vérifiée avant chaque extraction. N’utilisez jamais un flacon
endommagé.
Les flacons non gradués sont à usage unique. Les flacons gradués
peuvent être réutilisés tant qu’ils sont intacts.
5.25 cm
(2.07 in)
3.45 cm
(1.36 in)
Bottle Label
Okay
Vial
Label
Okay
Содержание ASE 150
Страница 1: ...ASE 150 Accelerated Solvent Extractor Operator s Manual Document No 065207 Revision 02 September 2008...
Страница 32: ...ASE 150 Operator s Manual 24 Doc 065207 02 9 08...
Страница 64: ...ASE 150 Operator s Manual 56 Doc 065207 02 9 08...
Страница 76: ...ASE 150 Operator s Manual 68 Doc 065207 02 9 08...
Страница 114: ...ASE 150 Operator s Manual 106 Doc 065207 02 9 08...
Страница 132: ...ASE 150 Operator s Manual 124 Doc 065207 02 9 08...