ASE 150 Operator’s Manual
12
Doc. 065207-02 9/08
Rinse Cells
Rinse cells are similar in appearance to sample cells, but are blue in color.
During a rinse cycle, solvent passes directly through the rinse cell and
into the collection vessel. For more information about rinse cycles, see
.
The rinse cell size must be matched to the size of the sample cell, as
indicated in the table below.
Ordering Sample Cells in Other Sizes
To perform a run with a cell in a different size than the size originally
ordered, order the cells from Dionex. You will also need to order the
Startup Kit appropriate for the new cell size. For cell and Startup Kit part
numbers, see
.
Note that installation of a different sample cell size will require a few
additional changes:
•
Reposition the cell holder to accommodate the new sample cell size,
if required. For instructions, refer to
•
Before beginning a run, specify the new cell size on the
SETUP
screen
Use this rinse cell:
With this sample cell:
Short (P/N 060174)
1 mL, 5 mL, 10 mL, 22 mL, 34 mL
Medium (P/N 060175)
66 mL
Long (P/N 060176)
100 mL
Содержание ASE 150
Страница 1: ...ASE 150 Accelerated Solvent Extractor Operator s Manual Document No 065207 Revision 02 September 2008...
Страница 32: ...ASE 150 Operator s Manual 24 Doc 065207 02 9 08...
Страница 64: ...ASE 150 Operator s Manual 56 Doc 065207 02 9 08...
Страница 76: ...ASE 150 Operator s Manual 68 Doc 065207 02 9 08...
Страница 114: ...ASE 150 Operator s Manual 106 Doc 065207 02 9 08...
Страница 132: ...ASE 150 Operator s Manual 124 Doc 065207 02 9 08...