ASE 150 Operator’s Manual
130
Doc. 065207-02 9/08
RINSE VOLUME
Specifies the volume of solvent rinsed through the cell
after the static heating step, expressed as a percentage of
the cell volume (0% to 150%). For example, if the
RINSE VOLUME
is 50%, 5 mL of solvent is rinsed
through a 10-mL cell, 17 mL of solvent is rinsed
through a 34-mL cell, and so on.
PURGE TIME
Specifies the amount of time the cell is purged with
nitrogen (0 to 900 seconds). Dionex recommends the
following settings:
•
40 to 80 seconds for a 1-mL, 5-mL, 10-mL, or
22-mL cell
•
70 to 110 seconds for a 34-mL cell
•
160 to 200 seconds for a 66-mL cell
•
250 to 290 seconds for a 100-mL cell
STATIC CYCLE
Specifies the number of times the static heating and
rinsing steps are performed (1 to 5). If more than one
cycle is specified, the rinse volume is divided among the
cycles.
CELL TYPE
Specifies the type of cell used:
SST
(stainless steel) or
Zr
(zirconium).
SAVE
Specifies the method number (1 through 24) under
which the selected parameters are saved. This field is
not available when a preprogrammed method is
selected.
Parameter
Description
Содержание ASE 150
Страница 1: ...ASE 150 Accelerated Solvent Extractor Operator s Manual Document No 065207 Revision 02 September 2008...
Страница 32: ...ASE 150 Operator s Manual 24 Doc 065207 02 9 08...
Страница 64: ...ASE 150 Operator s Manual 56 Doc 065207 02 9 08...
Страница 76: ...ASE 150 Operator s Manual 68 Doc 065207 02 9 08...
Страница 114: ...ASE 150 Operator s Manual 106 Doc 065207 02 9 08...
Страница 132: ...ASE 150 Operator s Manual 124 Doc 065207 02 9 08...