
B • Installation
Doc. 065207-02 9/08
119
Parameter
Description
METHOD
Specifies the name or number of the method to run.
CELL SIZE
Specifies which sample cell is installed (1 mL, 5 mL,
10 mL, 22 mL, 34 mL, 66 mL, or 100 mL). This setting
determines solvent volumes and temperature settings.
UNITS
Selects the unit of measure for pressure displays (psi,
MPa, bar, or atm).
REDUCE RELIEF
Determines whether the pressure relief valve is open
for a reduced period (
ON
) or the normal period (
OFF
).
When the pressure relief valve is open, residual
pressure is released from the sample cell. The default
setting (
OFF
) is recommended.
PREHEAT TIME
Specifies for how long the oven heats up without
solvent (0 to 99 minutes). The default setting (
0
) is
recommended.
PREHEAT PURGE
Enables and disables purging of the cell (without
solvent) during the preheat period. The default setting
(
OFF
) is recommended.
BYPASS HEAT-UP
Specifies whether to bypass the heat-up period for the
sample cell. The default setting (
OFF
) is recommended.
KEY SOUND
Determines whether a beep is emitted when a keypad
button is pushed.
ERROR SOUND
Determines whether a beep is emitted when an error
occurs (for example, you press the wrong keypad
button).
Table B-1. System Setup Parameters
Содержание ASE 150
Страница 1: ...ASE 150 Accelerated Solvent Extractor Operator s Manual Document No 065207 Revision 02 September 2008...
Страница 32: ...ASE 150 Operator s Manual 24 Doc 065207 02 9 08...
Страница 64: ...ASE 150 Operator s Manual 56 Doc 065207 02 9 08...
Страница 76: ...ASE 150 Operator s Manual 68 Doc 065207 02 9 08...
Страница 114: ...ASE 150 Operator s Manual 106 Doc 065207 02 9 08...
Страница 132: ...ASE 150 Operator s Manual 124 Doc 065207 02 9 08...