C • User Interface
Doc. 031842-01 5/02
C-7
C.1.4
Method Editor Screen
Use the
METHOD EDITOR
screen to define the parameters in a custom
method.
Figure C-7. Method Editor Screen (Additional text shown)
Parameter
Description
METHOD
Number of the method currently selected for editing.
TEMPERATURE
Temperature at which the extraction cell is heated (Off,
40 to 200 °C).
STATIC TIME
Static solvent extraction time (0 to 99 minutes).
FLUSH VOLUME
Amount of solvent flushed through the cell after the
static heating step, expressed as a percentage of the cell
volume (0 to 150%). For example, if the
FLUSH
VOLUME
is 50%, 5 mL of solvent is flushed through a
10 mL cell, 17 mL of solvent is flushed through a 34 mL
cell, and so on.
PURGE TIME
Amount of time the cell is purged with nitrogen (0 to
900 seconds). Dionex recommends the following
settings: 40 to 80 seconds for a 10 mL cell, 70 to 110
seconds for a 34 mL cell, 160 to 200 seconds for a 66
mL cell, and 250 to 290 seconds for a 100 mL cell.
STATIC CYLE
Number of times the static heating and flushing steps
are performed (1 to 5). If more than one cycle is
specified, the flush volume is divided among the cycles.
Содержание ASE 100
Страница 8: ...ASE 100 Accelerated Solvent Extractor vi Doc 031842 01 5 02...
Страница 24: ...ASE 100 Operator s Manual 2 10 Doc 031842 01 5 02 Figure 2 4 ASE 100 Rear Panel...
Страница 32: ...ASE 100 Operator s Manual 2 18 Doc 031842 01 5 02...
Страница 42: ...ASE 100 Operator s Manual 3 10 Doc 031842 01 5 02 Figure 3 2 Inspecting Cell O Rings...
Страница 60: ...ASE 100 Operator s Manual 3 28 Doc 031842 01 5 02...
Страница 69: ...4 Troubleshooting Doc 031842 01 5 02 4 9 Figure 4 1 ASE 100 Plumbing Diagram...
Страница 106: ...ASE 100 Operator s Manual 5 36 Doc 031842 01 5 02...
Страница 110: ...ASE 100 Operator s Manual A 4 Doc 031842 01 5 02...
Страница 113: ...B Installation Doc 031842 01 5 02 B 3 Figure B 1 ASE 100 Rear Panel...
Страница 150: ...ASE 100 Accelerated Solvent Extractor Index 8 Doc 031842 01 5 02...